background image

Inward Monitoring

Please check directly after delivery the device for any transport damages and deficiencies.
Additional with reference to the accompanying delivery note the number of parts must be
checked.
Claims for replacement or goods which relate to transport damage can only be considered
valid if the delivery company is notified without delay.

Functional description

The mechanical flow indicator DG06 is used for the optical control of flows and liquid
media. A stainless steel flap arranged in the flow chamber is lifted by the media flow.

Attention:

The DG06 is not a flow meter. It only indicates the presence of a flow. The numbers on
the scale cannot be assigned to a flow value. They only serve to indicate a certain flow
trend.

Maintenance and care

Due to the small number of moving parts, the units require very little maintenance. How-
ever, regular functional checks and maintenance not only increase the service life and
functional reliability of the device, but also that of the entire system.

The maintenance intervals depend on:

contamination of the medium

environmental conditions (e.g. vibrations)

DG06 Instruction manual  04/2019

page  3

Summary of Contents for DG04

Page 1: ...Instruction Manual DG06 Sight Flow Indicator with flap and scale PKP Prozessmesstechnik GmbH Borsigstraße 24 D 65205 Wiesbaden Nordenstadt Tel 49 0 6122 7055 0 Fax 49 0 6122 7055 50 Email info pkp de ...

Page 2: ...ce is prohibited and outside the scope of application In particular applications in which shock loads occur for example pulsed operation should be discussed and checked in advance with our technical staff The devices of the DG06 series should not be used as sole monitoring devices in order to detect or even avoid dangerous operating states in plants and machines The plant or machine itself must be...

Page 3: ...inless steel flap arranged in the flow chamber is lifted by the media flow Attention The DG06 is not a flow meter It only indicates the presence of a flow The numbers on the scale cannot be assigned to a flow value They only serve to indicate a certain flow trend Maintenance and care Due to the small number of moving parts the units require very little maintenance How ever regular functional check...

Page 4: ...ow chamber is lifted when there is flow and the current flow rate is displayed on the scale The flap is mounted on a stainless steel shaft it is raised by flow and lowered by gravity The DG06 can be installed both horizontally and vertically with flow from bottom to top and deployed in a wide variety of applications not least because it is not adversely affected by high tempera tures Typical appli...

Page 5: ...85 3 9 DN 20 3 4 95 140 89 66 1 85 3 9 DN 25 1 95 140 89 66 1 8 3 9 DN 32 1 1 4 120 180 120 89 4 7 1 DN 40 1 1 2 120 180 120 89 3 9 7 DN 50 2 150 220 170 110 9 14 5 DN 80 3 258 160 165 19 5 DN 100 4 258 160 165 23 5 G threaded F flanged Larger nominal sizes on request Order Code Order number DG06 Sight flow indicator with flap and scale S B G 15 0 Materials S cast steel E stainless steel Glass pre...