background image

V112712 

behind the fully racked in focuser.  The Secondary Spacer places the Ronchi screen of the 
Ronchi Ocular precisely where the focal plane is supposed to be. 
 
 

About the Ronchi Test

 

 
Figure 9 illustrates the basic idea behind 
using a Ronchi grating to identify the 
position of the focal plane. The light from 
a single bright star passes through the 
optical system and converges to a point on 
the focal plane. When the Ronchi grating 
is placed in front of or behind the focal 
plane, it will block some of the light in 
this light cone. This will cast a series of 
shadows that appear as dark, wide lines 
when the defocused light is viewed 
through the ocular.  
 

As the grating moves closer to the focal 
plane, fewer lines are visible, and they 
become wider. When the focal plane and 
the Ronchi screen are perfectly aligned, no lines (or perhaps one very wide line) are 
visible. See Figure 10 for an example. 
 
In order to set the primary-to-secondary distance, you will be moving the secondary 
mirror.  By adjusting secondary mirror distance, you will be moving the focal plane 
relative to the fixed position of the Ronchi screen.  The job here is to move the secondary 
mirror such that the focal plane coincides with the Ronchi screen on the Ronchi Ocular.  
When you have done this, you have accurately set the primary-to-secondary spacing.   
 
With reasonable effort, the primary-to-secondary spacing can be set to well within its 
1mm tolerance. 
 

Figure 9: The Ronchi grating casts fewer shadows as 
it moves closer to the focal plane. 

 

Summary of Contents for CDK Series

Page 1: ...V112712 1 PlaneWave CDK Telescope Instructions CDK12 5 17 20 and 24 ...

Page 2: ...optics must be collimated and the primary to secondary mirror spacing must be set Please note that the primary to secondary spacing is set at the factory and typically does not need to be adjusted unless it was moved after shipping Figure 1 The optical layout of the CDK system The system contains 4 elements the primary mirror the secondary mirror and a two element lens group Key Components Figure ...

Page 3: ...telescope at a bright star e Defocus the star until it becomes a donut like ring If the donut hole appears well centered in the donut proceed to Step 2 f If the donut hole is not centered adjust the collimating screws on the back of the secondary mirror so that the defocused star moves in the direction of the fat side of the donut as shown in Figure 4 Re center the star and repeat this process unt...

Page 4: ...ith the secondary preloaded securely against all knobs Collimating is a two person job one person looks through the eyepiece while one person adjusts the collimating screws Attempting to do this alone can be quite difficult and frustrating If necessary you might try using a webcam with a 1 1 4 nosepiece such as the Celestron NexImage in place of an eyepiece With this setup you can watch the video ...

Page 5: ...condary spacing is set at the factory You will typically not need to adjust this unless the secondary has been moved For your CDK optical system to perform as well as it should the spacing between the primary mirror and secondary mirror should be set to an accuracy of 1mm Fortunately you won t have to directly measure this spacing When the primary to secondary spacing is set correctly the focal pl...

Page 6: ...lar As the grating moves closer to the focal plane fewer lines are visible and they become wider When the focal plane and the Ronchi screen are perfectly aligned no lines or perhaps one very wide line are visible See Figure 10 for an example In order to set the primary to secondary distance you will be moving the secondary mirror By adjusting secondary mirror distance you will be moving the focal ...

Page 7: ...nes or one Ronchi line covers the entire aperture see Figure 10 this means that your spacing is already set you may proceed to Step 3 This is not uncommon because the primary to secondary spacing is set at PlaneWave and you most likely will not need to adjust it If you see two or more lines you should move the secondary toward or away from the primary The mechanics for moving the CDK12 5 CDK17 and...

Page 8: ...when the procedure is complete 3 Take a flat head screwdriver and place the tip in the central bolt of the secondary assembly Rotate the central bolt in a clockwise direction For reference rotating turn moves the secondary 4mm which moves the focal plane 3mm 4 Rotate the four collimation knobs in equal amounts until the assembly is tight again 5 Check the Ronchi screen and repeat this process as n...

Page 9: ...ounts For large spacing adjustments one may need to adjust the center knob shown in Figure 6 for the CDK20 or Figure 8 for the CDK24 To move the secondary toward the primary loosen the center knob and tighten the collimating screws by equal amounts until everything feels snug To move the secondary away from the primary loosen the three collimating screws by equal amounts and tighten the center kno...

Page 10: ...end of the altitude hub which is the mounting surface for the optical train Also the CDK700 is using the Focuser Rotator 600180 which uses 3 625 of backfocus on the CDK700 The CDK24 has significantly more backfocus than the other telescopes in the PlaneWave lineup In order to take advantage of the existing line of CCD spacers and accessories a spacer pushes the focuser out so it exactly matches th...

Page 11: ...re back focus than the other CDK models so a spacer is used to put the focuser the same distance from the focal plane as the other CDK models This means the same spacers will work with all the different models with exception of the CDK12 5 ...

Reviews: