background image

V112712 

 

Figure 10:  The left image is a simulation of what a null looks like through a Ronchi screen.  The right 
image is a simulation of what an out of focus view looks like through a Ronchi screen.  The right image 
shows 2 full lines.  The more lines you see, the farther the ronci screen is from the focal plane.   

 
 

Setting Up the Test 
 

Starting with a telescope that is fairly well collimated, center a bright star and replace the 
low power eyepiece with the Ronchi Ocular that was provided with the telescope.  Rack 
the focuser all the way in, as far as it goes. Make sure the 2” adapter is seated all the way 
into the focuser and the Secondary Spacer is seated flush into the 2” adapter.  If the 
Ronchi is nulled (you can see no lines, or one Ronchi line covers the entire aperture; see 
Figure 10), this means that your spacing is already set; you may proceed to Step 3.  This 
is not uncommon because the primary-to-secondary spacing is set at PlaneWave and you 
most likely will not need to adjust it.   
 
If you see two or more lines, you should move the secondary toward or away from the 
primary.  The mechanics for moving the CDK12.5/CDK17 and the CDK20 secondary 
mirrors are a little different.   
 

Determining the Direction to Move the Secondary (CDK12.5 / CDK17) 

 
To determine which way to move the secondary, gently slide the Ronchi Ocular out of 
the spacing tube.  If the Ronchi reaches null as you do this, the focal plane is too far back 
and you need to move the secondary mirror 

away

 from the primary.  If the number of 

lines grows as you slide the Ronchi out of the spacing tube, the focal plane is too far 
forward and you need to move the secondary 

toward

 the primary. 

 

Summary of Contents for CDK Series

Page 1: ...V112712 1 PlaneWave CDK Telescope Instructions CDK12 5 17 20 and 24 ...

Page 2: ...optics must be collimated and the primary to secondary mirror spacing must be set Please note that the primary to secondary spacing is set at the factory and typically does not need to be adjusted unless it was moved after shipping Figure 1 The optical layout of the CDK system The system contains 4 elements the primary mirror the secondary mirror and a two element lens group Key Components Figure ...

Page 3: ...telescope at a bright star e Defocus the star until it becomes a donut like ring If the donut hole appears well centered in the donut proceed to Step 2 f If the donut hole is not centered adjust the collimating screws on the back of the secondary mirror so that the defocused star moves in the direction of the fat side of the donut as shown in Figure 4 Re center the star and repeat this process unt...

Page 4: ...ith the secondary preloaded securely against all knobs Collimating is a two person job one person looks through the eyepiece while one person adjusts the collimating screws Attempting to do this alone can be quite difficult and frustrating If necessary you might try using a webcam with a 1 1 4 nosepiece such as the Celestron NexImage in place of an eyepiece With this setup you can watch the video ...

Page 5: ...condary spacing is set at the factory You will typically not need to adjust this unless the secondary has been moved For your CDK optical system to perform as well as it should the spacing between the primary mirror and secondary mirror should be set to an accuracy of 1mm Fortunately you won t have to directly measure this spacing When the primary to secondary spacing is set correctly the focal pl...

Page 6: ...lar As the grating moves closer to the focal plane fewer lines are visible and they become wider When the focal plane and the Ronchi screen are perfectly aligned no lines or perhaps one very wide line are visible See Figure 10 for an example In order to set the primary to secondary distance you will be moving the secondary mirror By adjusting secondary mirror distance you will be moving the focal ...

Page 7: ...nes or one Ronchi line covers the entire aperture see Figure 10 this means that your spacing is already set you may proceed to Step 3 This is not uncommon because the primary to secondary spacing is set at PlaneWave and you most likely will not need to adjust it If you see two or more lines you should move the secondary toward or away from the primary The mechanics for moving the CDK12 5 CDK17 and...

Page 8: ...when the procedure is complete 3 Take a flat head screwdriver and place the tip in the central bolt of the secondary assembly Rotate the central bolt in a clockwise direction For reference rotating turn moves the secondary 4mm which moves the focal plane 3mm 4 Rotate the four collimation knobs in equal amounts until the assembly is tight again 5 Check the Ronchi screen and repeat this process as n...

Page 9: ...ounts For large spacing adjustments one may need to adjust the center knob shown in Figure 6 for the CDK20 or Figure 8 for the CDK24 To move the secondary toward the primary loosen the center knob and tighten the collimating screws by equal amounts until everything feels snug To move the secondary away from the primary loosen the three collimating screws by equal amounts and tighten the center kno...

Page 10: ...end of the altitude hub which is the mounting surface for the optical train Also the CDK700 is using the Focuser Rotator 600180 which uses 3 625 of backfocus on the CDK700 The CDK24 has significantly more backfocus than the other telescopes in the PlaneWave lineup In order to take advantage of the existing line of CCD spacers and accessories a spacer pushes the focuser out so it exactly matches th...

Page 11: ...re back focus than the other CDK models so a spacer is used to put the focuser the same distance from the focal plane as the other CDK models This means the same spacers will work with all the different models with exception of the CDK12 5 ...

Reviews: