background image

V112907 

8

 

Figure 8:  The left image is a simulation of what a null looks like through a Ronchi screen.  The right image 
is a simulation of what an out of focus view looks like through a Ronchi screen.  The right image shows 2 
full lines.  The more lines you see, the farther the ronci screen is from the focal plane.   

 
After you are satisfied that the secondary spacing is correct, replace the Ronchi ocular 
with the low power eyepiece and repeat step 1 to verify that the collimation is still close.  
If re-collimation adjustments are necessary, make them and repeat step 2. 

 

Step 3, Fine Collimation: 

 

Use a high power eyepiece (5mm or less) and follow the same procedure described in 
Step 1 (adjust the secondary so that the donut hole is in the center of the donut).  You 
may want to switch to low power when making adjustments and re-centering the star and 
back to high power to check the collimation.  Finally, use the Ronchi eyepiece to re-
check the spacing.  As long as only small adjustments were made in Step 3, the spacing 
should still be fine.  If it is not, repeat Step 2, and Step 3. 
 
 
 
 
 
 

Summary of Contents for CDK

Page 1: ...V112907 1 PlaneWave CDK Collimation Instructions ...

Page 2: ...d secondary To get the best performance out of the CDK optical system the optics must be collimated and secondary to primary mirror spacing must be set Figure 1 The optical layout of the CDK system The system contains 4 elements which consist of the primary mirror the secondary mirror and a 2 element lens group Key Components Figure 2 Here are the standard accessories used for collimation and sett...

Page 3: ...e telescope at a bright star Defocus the star until it becomes a donut looking ring If the donut hole looks centered in the center of the donut proceed to Step 2 If the donut hole is not centered adjust the collimating screws on the back of the secondary mirror so that the defocused star moves in the direction of the fat side of the donut as shown in figure 4 Re center the star and repeat this pro...

Page 4: ... to each other Rotate one knob clockwise and one knob counter clockwise in order to keep the tension between all the knobs Always finish a move and with all the secondary preloaded securely with the knobs Collimating is a two person job One person at the eyepiece while one person adjusts the collimating screws Attempting to do this alone can be quite difficult and frustrating Figure 6 CDK20 second...

Page 5: ...ides with the Ronchi screen on the Ronchi Ocular When you have done this you have accurately set the primary to secondary spacing With reasonable effort the primary to secondary spacing can be set to well within its 1mm tolerance Now that the collimation is at least close and a bright star is centered replace the low power eyepiece with the provided Ronchi Ocular Rack the focuser all the way in as...

Page 6: ... complete 3 Take a flat head screw driver and place the tip in the central bolt of the secondary assembly Rotate the central bolt in a counter clockwise direction For reference rotating turn moves the secondary 4mm and that moves the focus 3mm 4 Take the four collimation knobs and rotate them in equal amounts until the assembly is tight again 5 Check the Ronchi screen and repeat process as necessa...

Page 7: ...ermine which way to move the secondary gently slide the Ronchi ocular out of the spacing tube If as you do this the Ronchi reaches null the focal plane is too far back and you need to move the secondary mirror away from the primary If as you slide the ronchi out of the spacing tube the number of lines grows the focal plane is too far forward and you need to move the secondary toward the primary Fo...

Page 8: ...peat step 1 to verify that the collimation is still close If re collimation adjustments are necessary make them and repeat step 2 Step 3 Fine Collimation Use a high power eyepiece 5mm or less and follow the same procedure described in Step 1 adjust the secondary so that the donut hole is in the center of the donut You may want to switch to low power when making adjustments and re centering the sta...

Reviews: