PLANMECA INTRA X-RAY UNIT 

MOBILE BASE

 INSTALLATION INSTRUCTIONS

1.

Attach the generator back plate to the column with eight M6x16 DIN 912 screws (1) using the 5mm
Allen key. Attach the handle to the generator back plate with two M10x40 DIN 912 screws (2) using the
8mm Allen key.

mobile_intr

a1.eps

1

2

Column

Generator back plate

Handle

COPYRIGHT PLANMECA 2004-05
PUBLICATION NUMBER 10006705 version 3

Summary of Contents for INTRA X-RAY UNIT

Page 1: ...e to the column with eight M6x16 DIN 912 screws 1 using the 5mm Allen key Attach the handle to the generator back plate with two M10x40 DIN 912 screws 2 using the 8mm Allen key mobile_intra1 eps 1 2 Column Generator back plate Handle COPYRIGHT PLANMECA 2004 05 PUBLICATION NUMBER 10006705 version 3 ...

Page 2: ...base plate can be attached to the column base plate s and column s screw holes in line 3 Make sure that the gap between the column edge and column holder is approx 3mm 0 12 in so that the column is firmly attached to the base plate Attach the base plate to the col umn with four M10x40 DIN 912 screws 4 using the 8mm Allen key Lift the mobile base up mobile_intra2 eps 1 3 2 4 2 3 mm Column Column ho...

Page 3: ... below Route the Intra cables through the horizontal arm and column Attach the horizontal arm to the column with two attachment screws 1 and larger 2 and smaller 3 washers 1 2 3 mobile_intra7 eps 5x18 pin Pin 5x14 pin 5x18 pin 5x14 pin 5x14 pin 5x14 pin Gap Place the cylindrical pins to the bearing housing according to figure beside Horizontal arm Intra cables ...

Page 4: ...e bracket arm with the locking plate 5 Carefully pull the Intra cables out from the column so that they can be connected to the generator PCB Leave the extra length of the cables into the column Attach the generator assembly to the generator back plate with four attachment screws mobile_intra8 eps Locking plate M5x6 DIN 912 screw Use 4mm Allen key mobile_intra5 eps Generator assembly Intra cables ...

Page 5: ...ion instructions PLANMECA INTRA X RAY UNIT 6 Connect the Intra cables to the Generator PCB according to the figure below 7 Attach the generator housing to the generator assembly with three M4x6 ULS screws mobile_intra P1 P8 mobile_intra11 eps ...

Page 6: ...s horizontal arm and column top Insert the plugs into the attachment screw openings on the column see figure below Secure the bracket arm to the transport position with the strap supplied with the base CAUTION NEVER MOVE THE INTRA MOBILE X RAY UNIT WITHOUT FIRST SECURING THE BRACKET ARM TO THE TRANSPORT POSITION mobile_intra10 eps Strap ...

Page 7: ...e power supply and hang the power supply cable on the handle Transport the x ray unit with care Do not bump the unit down steps or allow the tube head to hit anything The mobile base has mechanical locking levers on the back wheels Press the locking levers down to lock the wheels This will prevent the mobile base from moving If you have to move the mobile base over a low obstacle tilt the base sli...

Page 8: ...PLANMECA INTRA X RAY UNIT 8 8 Planmeca Intra Mobile base Installation instructions ...