6 MAIN PARTS
14 Planmeca ProMax 2D & 3D s & 3D Classic
User’s Manual (2D)
6.8 Touch screen
NOTE
The options shown on the touch screen depend on the unit
configuration. The X-ray unit can be upgraded with new
programs and features, contact your dealer for further
information. The views and values shown in this manual
are only examples.
NOTE
The illustrations shown on the touch screen are based on
approximate patient anatomy. The actual exposure area
depends on the individual anatomy of the patient.
NOTE
Never allow patients to touch the screen when they are
positioned in the X-ray unit. Touching the screen during
exposure will stop the imaging process.
You can use the buttons at the bottom of the main screen
to change the appearance of the main view.
•
Left button:
To view the ProMax model and up to five most
recently used programs, select the left button at the
bottom of the main screen. The most recently used
program is shown first. This is the default view of the
main screen.
NOTE
If you wish to use fast forward buttons on this view,
select Settings > Program > 2200 Program Features
> Fast Forward ON. Using a fast forward button takes
you directly to the last screen where you can take an
exposure.
•
Right button:
To view two program bars on top of each other,
select the right button at the bottom of the main
screen.
Summary of Contents for Planmeca ProMax 3D Classic
Page 1: ...PlanmecaProMax 2D 3D s 3D Classic EN 10033256_8 user s manual 2D imaging ...
Page 2: ......
Page 88: ......
Page 89: ......