Note: This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and,
if not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area is likely
to cause harmful interference in which case the user will be required to correct the interference
at his own expense.
This ISM device complies with Canadian ICES-001. (Cet appareil ISM est conforme à la norme
NMB-001 du Canada.)
Warning: This is a Class A product. In a domestic environment this product may cause radio
interference in which case the user may be required to take adequate measures.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The PlanScan is intended for use in the electromagnetic environment specified below. The
customer or the user of the PlanScan should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment -
guidance
RF emissions CISPR 11
Group 1
The PlanScan uses RF energy only for
its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11
Class A
The PlanScan is suitable for use in all
establishments other than domestic
and those directly connected to the
public low voltage power supply
network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC
61000-3-2
Not
applicable
Voltage fluctuations
/ flicker emissions IEC
61000-3-3
Not
applicable
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The PlanScan is intended for use in the electromagnetic environment specified below. The
customer or the user of the PlanScan should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV
contact
+/- 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast
transient / burst
IEC 61000-4-4
+/- 2 kV for power
supply lines
+/- 1 kV for input/
output
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment
10828200.B
Scanner Specifications
49
Summary of Contents for PlanScan
Page 1: ...Planmeca PlanScan User Manual ...
Page 2: ......