76
5. Discard the first drop of Control Solution and squeeze a small
drop of Control Solution on a clean nonabsorbent surface (such
as a clean piece of wax paper). Do not apply Control Solution to
the Test Strip directly from the bottle.
6. Bring the tip of the Test Strip to lightly touch the drop of Control
Solution. Control Solution is automatically pulled into the strip
through the tip. Hold until the meter beeps. The meter will now
begin counting down from 5 to 1 and the Control Solution test
result will appear.
CAUTION:
Please remember that in order to prevent contamination, follow the above
instructions for performing the Control Solution test.