3.3.2
Tare function
When switching on the tare function, the instantaneous measured value is set to
zero. In this case, measurement is continued with this offset. By switching off the
tare function, the actual measured value is displayed again.
measured value
effective value
Tare can be activated during configuration (
Func
r
Fnc.1
= 1).
Dependent of configuration, tare can be made effective via one of the digital in-
puts di1, di2, di3, the
è
key or interface (
LOGI
r
tArA
).
3.3.3
Sample&hold amplifier
With the sample & hold function activated, the measured value is held on the dis-
play. By switching off the sample & hold function, the actual measured value is
displayed again.
measured value
effective value
Sample & hold can be activated during configuration (
Func
r
Fnc.1
= 2).
Dependent of configuration, sample & hold can be made effective via one of the
digital inputs di1, di2, di3,the
è
-key or via interface (
LOGI
r
HOLd
).
Operation
Digital 280-1