2-point correction
(
ConF
/
InP.1
/
Corr
=
1
):
InL.1
:
The input value of the lower scaling point is displayed.
The operator must adjust the lower input value by means of a
process value simulator and confirm the input value by pressing key
Ù
.
OuL.1
:
The display value of the lower scaling point is displayed.
Before calibration,
OuL.1
equals
InL.1
.
The operator can correct the lower display value by pressing the
ÈÌ
keys. Subsequently, he confirms the display value by pressing key
Ù
.
InH.1
:
The input value of the upper scaling point is displayed. .
The operator must adjust the upper input value by means of the
process value simulator and confirm the input value by pressing key
Ù
.
OuH.1
:
The display value of the upper scaling point is displayed.
Before calibration
OuH.1
equals
InH.1
.
The operator can correct the upper display value by pressing keys
ÈÌ
Subsequently, he confirms the display value by pressing key
Ù
.
Calibration level
Digital 280-1
Offset correction
39
r
Ì
Ù
3 sec.
r
PArA
CAL
r Ù r
InP
Ù
r
r
r Ù
End
r
r Ù
InL
È
InL
Ù
OuL
r Ù
Ì
È
r Ù
InH
È
InH
Ù
OuH
r Ù
Ì
È
FUNC
ADA
OUT1
OUT2
OUT3
ERR
Digital 280-1
ConF
r
Ì
r
Summary of Contents for Digital 280-1
Page 48: ...12 Notes Notes Digital 280 1 48...
Page 49: ...Notes 49 Digital 280 1...
Page 50: ...Digital 280 1 50...