Cleaning Test Platform DAILY
• To remove, open the cartridge door by
pressing the button located on the lower
right side of meter.
• Lift up and pull from the end closest to
the cartridge door.
• Using a Clorox Germicidal Wipe, clean
both front and back of test port,
including INSIDE test port hole.
Disinfecting Test Platform DAILY
Always make sure any visible blood
and debris is cleaned off test platform
before handling
(See Cleaning Test Platform DAILY above)
.
• Using a
NEW
Clorox Germicidal Wipe,
thoroughly wipe front and back surfaces.
Disinfect the test port hole by pushing
wipe through.
• Allow the test platform surface to remain
damp for one minute, then let air dry
completely.
• Leave off until meter interior is
disinfected.
Test Port Hole
Test Port
Test Platform
Disinfect both front and
back surfaces of test
platform
REMEMBER: Disinfect
inside test port hole
CAUTION
: Handling the test port hole after
disinfecting may recontaminate it.
DO NOT
TOUCH
while placing back on meter.
• To replace, grab test platform by either
bottom corner, insert top tab into open
slot in meter.
• Snap bottom two tabs into place.
• Ensure platform is completely in place
before testing.