background image

Remove Batteries

Customer Service 1-855-IMI-POGO (464-7646) • www.presspogo.com

71

70

Customer Service 1-855-IMI-POGO (464-7646) • www.presspogo.com

IMPORTANT:

 If you have questions about cleaning or 

disinfecting your POGO meter, please call Intuity Medical 
Customer Service at 1-855-IMI-POGO (464-7646). If you notice 
signs of deterioration, discontinue use and call Intuity Medical 
Customer Service to find out about replacement. Examples of 
meter damage or malfunction include: fogged screen, cracked 
screen or case, illegible labels, buttons not working, failure to 
move to the next test, cartridge door does not open, meter fails 
to test, test port appears cracked or broken, repeated error 
messages or multiple failed control solution tests.

*The expected use life of the meter is three years. 

Meter and Meter Parts Three-Year Life Cycle*

Replacing the Batteries

Your POGO meter comes with two pre-
installed AAA batteries. The batteries 
should last for approximately 500 tests. 
When the battery power runs low, the  
low battery symbol appears in the status 
bar. When this occurs, the meter has 
enough power for about 50 tests. Accuracy 
will not be affected. However, you should 
change your batteries as soon as possible. 

Removing the batteries will not affect 
the meter memory or previously stored 
settings. The time and date will be restored 
to the last time and date displayed before 
battery removal.

IMPORTANT:

 When the dead battery symbol appears, your 

POGO meter will not perform a test. Change the batteries

 

immediately.  

•  Remove battery door on back of meter  

by sliding the door toward the top end  
of the meter. 

•  Remove and discard the used alkaline 

batteries according to your local 
environmental regulations. 

•  Install two new AAA batteries side-by-

side into the battery compartment. Be 
sure to align the positive and negative 
ends as labeled inside the compartment. 

Re-insert Batteries

Low Battery

Dead Battery

Care 

Example Usage

Cycles 

Test Platform Daily Cleaning & 
Disinfection

365 days x 3 yrs

1095 Cycles

Total Tested

1100 Cycles

Meter Cleaning & Disinfection After 
Every Cartridge or Weekly

Every 1.5 days  

x 3 yrs

730 Cycles

Total Tested

730 Cycles

IMPORTANT SAFETY DOCUMENTS FOR YOUR REFERENCE

“FDA Public Health Notification: Use of Fingerstick Devices on More than One 
Person Poses Risk for Transmitting Bloodborne Pathogens: Initial Communication” 
(2010) https://www.fda.gov/MedicalDevices/ProductsandMedicalProcedures/
InVitroDiagnostics/ucm227935.htm

“CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person 
Poses Risk for Transmitting Bloodborne Pathogens” (2010) http://www.cdc.gov/
injectionsafety/Fingerstick-DevicesBGM.html

C

A

R

IN

G

 F

O

R

 YO

U

R

 

POG

O

C

A

R

IN

G

 F

O

R

 YO
U

R

 

POG

O

Summary of Contents for Press. Test. GO

Page 1: ...POGO Automatic Blood Glucose Monitoring System Owner s Manual Press Test GO Simply press finger here for easy One Step testing For Single Patient Use Only...

Page 2: ...e hope you enjoy using your POGO Automatic System Indications For Use The POGO Automatic Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose sugar in fre...

Page 3: ...immediately DO NOT use the meter to test your blood glucose until you obtain a passing result This device is not intended for use in healthcare or assisted use settings such as hospitals physician off...

Page 4: ...de significant health benefits for people with diabetes The Diabetes Control and Complications Trial DCCT showed that tight glucose control can reduce the risk of diabetes related complications such a...

Page 5: ...s Appear Instead of Result 40 Assisting Patients with Automatic Testing 41 Pre or Post Meal Marking optional 44 Understanding Your Results 45 Blood Glucose Levels 46 Troubleshooting Tips 47 How Do You...

Page 6: ...iters 1 6 3 25 Other Meter Brands POGO Not actual size The POGO Automatic Meter Time Date USB data port Bluetooth Connectivity see page 52 Tests remaining in cartridge Display screen Review test resul...

Page 7: ...has enough power for about 50 tests Accuracy will not be affected However change the battery as soon as possible Large low battery symbol displayed indicates battery is dead Bluetooth Date Time Result...

Page 8: ...artridge foil cup was opened The meter will display this as a number Meter temperature is out of operating temperature range Remove cartridge This is displayed along with an alert error code See Troub...

Page 9: ...Results Settings Menu 2 15 PM 2 10 2019 12h 12 24 ON Code JZSK Apple iPhone 8 Google Pixel Samsung S8 Add Device Note Always remember to set time and date Time Date Settings Quick Graph Settings Sele...

Page 10: ...sealed and contains one test strip and one sterile lancet Test cartridges are single patient use only Do not share cartridges even with family members Each lancet is 30 gauge Single use only and auto...

Page 11: ...the next available test Do not share partially used cartridges even with family members Eject into hand when removing a partially used cartridge for later use Handle with care When grasping partially...

Page 12: ...Cartridge Open the cartridge door by pressing the button on the lower right side of the meter Insert a new test cartridge into the meter with the bottom facing inside the meter The cartridge fits int...

Page 13: ...s will find it sufficient to use Level 2 Applicator handle POGO Level 1 Control Solution Applicator POGO Level 2 Control Solution Applicator When Should You Perform a Control Solution Test To practice...

Page 14: ...did not receive enough control solution from the applicator You have 90 seconds to complete the control test When there are 60 seconds left the meter will count down to 0 Pick up the control solution...

Page 15: ...ters of blood The POGOAutomatic System is a new way of testing POGO Automatic has All Automatic technology which means lancing and blood collection are combined in one easy step Just place your finger...

Page 16: ...gertips are at room temperature or warmer before testing IMPORTANT If you have difficulty getting enough blood CHOOSE THE RIGHT TEST SITE AND INCREASE CIRCULATION TO FINGERTIP PRESS DOWN FIRMLY on tes...

Page 17: ...port hole Continue pressing down to start 3 2 1 countdown At the end of the countdown you will be lanced YOU MUST HOLD FINGER IN PLACE after lance IMPORTANT If you lift your finger after lance it will...

Page 18: ...r collects your blood sample IMPORTANT DO NOT PLACE FINGER ON TEST PORT until meter signals ready to test It may stop cartridge from rotating or cause meter malfunction or damage If this occurs a CC m...

Page 19: ...during testing your results are displayed as a redo the test If this continues call Intuity Medical Customer Service at 1 855 IMI POGO 464 7646 Glucose result Quick Graph In Range Out of Range Low Ou...

Page 20: ...INGER from test port and gently squeeze to get more blood Add More Blood Guidance Keep adding blood until enough for result IMPORTANT Continue to add blood as many times as needed within the 90 second...

Page 21: ...e sure to wear gloves and follow the meter display symbols See Testing Symbols on page 15 for complete POGO test process Assisting Patients with Automatic Testing IMPORTANT If the meter is being opera...

Page 22: ...over test port hole Center and PRESS DOWN into test cell HOLD until meter indicates Remove patient s finger from test port and squeeze lanced area to get more blood Center the blood droplet over the...

Page 23: ...e correct time and date on your meter before testing If incorrect your mealtime averages will not adequately reflect your glucose patterns see Setting Up Your Meter on page 55 CAUTION LO low and HI hi...

Page 24: ...re professional and follow their treatment advice 3 If condition worsens seek medical attention immediately Never make significant changes to your treatment regimen without consulting your healthcare...

Page 25: ...comparing your lab result with your POGO Meter result Recalling Results from Memory In memory mode you can view the last 500 test results as well as result averages The symbol appears indicating you a...

Page 26: ...access your averages press the M button again see diagram below while in Memory Mode Your averages appear along with the number of results in each average n of results in average Example Blood Glucose...

Page 27: ...tterns app on your smartphone You should also have Bluetooth enabled ON the phone The POGO meter will automatically turn on Bluetooth when you start the pairing process To pair your POGO use the follo...

Page 28: ...results can be recorded with the correct information and averages can be correctly calculated To change the time make sure the blinking selection arrow is next to the clock symbol and press M to sele...

Page 29: ...ll buttons to select the back arrow at bottom of screen Press the M button to confirm Set Up Settings Select the Set Up Settings Menu from the Main Set Up Menu see page 54 Sound The meter uses sounds...

Page 30: ...you prefer Use the scroll buttons to either turn the vacuum ON or OFF Press the M button to confirm setting Use the back arrow at bottom of screen to return to the Main Set Up Menu Bluetooth Auto Time...

Page 31: ...e Meter Pairing Symbol Press the M button to enter the pairing screen Follow the instructions on Page 52 to pair your POGO meter with your phone Deleting a Paired Device from POGO To remove a device u...

Page 32: ...imen for removing dirt and debris prior to disinfection It does not kill germs Disinfection is the only way to lessen your exposure to infectious diseases through blood transmittal If blood is visible...

Page 33: ...Cleaning Test Platform DAILY above Using a NEW Clorox Germicidal Wipe thoroughly wipe front and back surfaces Disinfect the test port hole by pushing wipe through Allow the test platform surface to re...

Page 34: ...micidal Wipes as needed to clean ALL meter surfaces Clean meter exterior Wipe front back sides and buttons of meter case Clean meter interior Thoroughly wipe all interior walls and corners inside cart...

Page 35: ...eeds to be disinfected using a NEW Clorox Germicidal Wipe Be sure to disinfect AFTER EVERY CARTRIDGE OR WEEKLY whichever comes first To clean see Cleaning Meter on page 66 Disinfect all interior meter...

Page 36: ...affect the meter memory or previously stored settings The time and date will be restored to the last time and date displayed before battery removal IMPORTANT When the dead battery symbol appears your...

Page 37: ...compartment Replace Battery Door NOTE Be sure to set the correct time and date on your meter before testing see Setting Up Your Meter on page 55 Storage and Handling Store POGO at room temperature be...

Page 38: ...r 2 Incomplete Test meter error E 2 Meter could not calculate blood glucose result Meter did not read strip Repeat test No Blood Alert NO No blood collected on test strip Repeat test after rubbing you...

Page 39: ...Try testing with a new cartridge Error 6 Barcode Error E 6 Meter could not read barcode data from cartridge Re insert cartridge If error repeats insert new cartridge Error 1 System Failure E 1 A prob...

Page 40: ...tronic device you should follow all local environmental regulations for disposal Bluetooth Wireless Technology Your POGO Automatic meter uses Bluetooth low energy technology to communicate and transfe...

Page 41: ...r equivalent product free of charge The warranty on the replacement will expire on the date of the original warranty expiration or 90 days after the shipment of a replacement system whichever period i...

Page 42: ...ITTED BY APPLICABLE LAW OUR MAXIMUM AGGREGATE LIABILITY TO YOU OR TO ANY THIRD PARTY CLAIMING THROUGH OR UNDER YOU IN DAMAGES OR OTHERWISE SHALL NOT EXCEED THE PURCHASE PRICE PAID TO US BY YOU FOR THE...

Page 43: ...500 West Warren Avenue Fremont CA 94538 1 855 IMI POGO 464 7646 www presspogo com 2019 Intuity Medical Inc All Rights Reserved ART 02967 Rev 03 Patent information available at www presspogo com patent...

Reviews: