SprintScan 35LE Service Manual
Disassembly
1 6
Removing Scanner Fuses
Notes:
•
If the POWER indicator, READY indicator and scanner lamp all fail to
light up when the SprintScan 35LE scanner is turned on, first make sure
that power exists at the source.
•
If power availability can be confirmed and the scanner’s lights still fail to
come on, check the fuse.
There are two fuses installed in the SprintScan 35LE scanner: one is located at the rear
of the scanner and the other is located in the power supply.
Replacing System Fuse
1. Turn off the power and disconnect the power cord.
2. Use a flat screw driver to lever over the fuse lid (Figure 2-12).
Figure 2-12. Replacing system fuse
3. Replace the fuse. (A spare fuse is located close to the rear side.)
Fuse Specifications (At the rear)
UL and CSA
(U.S.A)
Bel 5MF 1.6, Buss GMA-1 6/10,
Little 23501.6A
TUV GS
(Europe)
Bel 5SF 1.6, Eska 520.600,
Little 2170.16, Olvis EA,
Schurter SP001.1006
Fuse