28
POLITEC | INSTALLATION MANUAL - VER.3.1
RX 1
TX 1
RX 2
TX 2
RX 3
TX 3
RX 4
TX 4
N.B.:During the alignment test phase, the
activation of a TX transmitter determines the
automatic shutdown of the other TX optics in
the column.
16.3 Parallel beam calibration
Test the TX1 and RX1 optics and proceed with the calibration as explained.Then repeat the operation
for each pair of optics.
16.4 Calibration with active crossing function
To activate the function, move DIP 2 of the 12 DIPSWITCH unit of the MES9012 in ON.
RX 1
TX 1
RX 2
TX 2
RX 3
TX 3
RX 4
TX 4
ON
1
2 3 4 5 6 7 8 9 10 11 12
Test the TX1 and RX1 optics and proceed with the calibration as explained above but, having found the
maximum alignment, keep the TX1 optic in test and exit the test only with RX1, then repeat the same
operations with RX2, RX3 and RX4.
At the end, exit the test of TX1 and repeat the operation for each TX optic on each RX.
At the end of all operations, make sure that TX1 and TX4, being at the ends and most susceptible to
orientation movements, are still optimally aligned with the corresponding RX1 and RX4.
MANA DT SMA
WARNING
: too close a distance
between the TX and RX columns does
not allow calibration with the crossing
function.