background image

                                                                                                                        3025

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

                      Figure 23 

Adjusting the Aperture Diaphragm 

The  aperture  diaphragm  is  designed  for  the  adjustment  of  the  numerical  aperture,  not  for  brightness. 

Generally,  reducing  the  diaphragm  opening  to  70-80%  of  the  N.A.  value  of  the  respective  objective 

will  provide  an  image  of  acceptable  quality.  If  you  want  to  observe  the  image  of  the  aperture 

diaphragm, remove one eyepiece and look through the tube. You will see a dark circle encroaching on 

the bottom of the tube 

 

 

 

                                 

Swing Out Condenser 

Aperture Diaphragm 

 

Figure 21 

 

Figure 22 

4-7 Focusing (Figures 21 & 22) 

 

Push the light path selector lever (Figure 25) completely in. 

Lower  the  mechanical  stage.  With  the  10x  objective  in 

position, raise the mechanical stage slowly using the coarse 

focusing  knob  until  the  specimen  is  in  focus.  Then  rotate 

the fine focus until a sharp image is obtained. Do not allow 

the objective to touch the specimen.   

 

 

 

4

44

4-

--

-8

88

  

 

Adjusting  the  Swing  Out  Condenser 

(Figure 23) 

The  center  of  the  condenser  and  the  light  axis  of  the 

objective are coaxial. They have been factory adjusted   

and  does  not  need  to  be  adjusted  .  The  upper  limit  of 

the condenser has also been adjusted.   

 

Turn  the  condenser  focus  knob  to  adjust  the 
condenser.  Raise  the  condenser  when  using  the 
higher  magnification  objectives,  and  lower  the 
condenser  when  using  lower  magnification 
objectives. 
 

Swing  out  the  condenser  and  away  from  the 

light  path  when  using  the  4x  objective.  Swing   

in  the  condenser  and  into  the  light  path  when 

using  the  10x  and  higher  magnification 

objectives. 

13

Summary of Contents for 3025 series

Page 1: ...wan 10595 886 2 8712 0600 886 2 8712 2677 fax info polysciences tw Polysciences Inc 400 Valley Road Warrington PA 18976 800 523 2575 215 343 6484 800 343 3291 215 343 0214 fax info polysciences com ww...

Page 2: ...y notes i Care and Maintenance ii 1 Components Illustration 1 2 Installation of Components 4 2 1 Installation Diagram 4 2 2 Installation Procedures 5 3 Operation 11 4 Technical Specifications 15 5 Tro...

Page 3: ...and wipe up the spillage Otherwise the instrument may be damaged 5 CAUTION the lamp lamp housing and adjacent parts will become very hot Do not touch these parts until they have completely cooled Nev...

Page 4: ......

Page 5: ...1 COMPONENTS ILLUSTRATION...

Page 6: ...2...

Page 7: ...Microscope Body Lamp House Mechanical Stage Mounting Assembly 10 Wide Field Eyepiece Infinity Plan Objectives Trinocular Viewing Head Video Port optional 3...

Page 8: ...Figure 2 2 2 Assembly 3025 4 Guide Board Figure 3 Locking Block and Bolt Figure 1 4...

Page 9: ...ench to connect the mechanical stage mounting support assembly and the guide board Ensure the connection is tight and secure 2 Place the mechanical stage mechanical assembly Figure 2 onto the top of t...

Page 10: ...g the Lamp Figure 6 Halogen Lamp 6 Volt 30 Watt Do not use a lamp of a different voltage or wattage 1 Turn the power switch to the off position and remove the power cord 2 Allow the lamp house and bul...

Page 11: ...l the lowest magnification objective into the nosepiece Fig 9 Then in a clock wise direction rotate the nosepiece and install the objectives in succeeding higher magnification sequence Fig 10 Inspect...

Page 12: ...ive caps from the eyepiece tubes Insert the eyepieces into the eyepiece tubes Fig 11 2 2 7Installing the Video Port optional Remove the protective cap from the vertical tube of the viewing head Insert...

Page 13: ...Upper Stage Limit Control Fine Focus Knob Coarse Focus Knob Field Diaphragm Condenser Adjustment Knob 9...

Page 14: ...ng Out Condenser with Iris Diaphragm Tension Control Collar Y Stage Movement Adjustment Knob X Stage Movement Adjustment Knob Interpupillary Distance Indicator Variable Light Intensity Adjustment Dial...

Page 15: ...able intensity dial Use of the lamp at a lower intensity will prolong its life 4 3Adjusting Tension Adjustment Collar Figure 17 The tightness of the tension adjustment collar is factory adjusted If th...

Page 16: ...tance range 48mm 75mm While observing with two eyes hold the left and right eyetubes Rotate the eyetubes around the central axis Adjust the interpupillary distance until the left and right fields of v...

Page 17: ...ve in position raise the mechanical stage slowly using the coarse focusing knob until the specimen is in focus Then rotate the fine focus until a sharp image is obtained Do not allow the objective to...

Page 18: ...14...

Page 19: ...ut condenser N A 0 9 Configuration Table Viewing Head Compensation Free Trinocular Head Eyepiece Extra Wide Field Eyepieces EW10x 22 Objectives Infinite Plan objectives 4x 10x 40x 100x Oil Infinite Pl...

Page 20: ...16...

Page 21: ...pside down Turn the slide over so the high power objectives cover glass faces up Cover glass is to thick Use a 0 17mm cover glass High power objective contacts Slide upside down Turn the slide over so...

Page 22: ...Dual Observation 1 unit Order online anytime at www polysciences com Polysciences Europe GmbH Handelsstrasse 3 D 69214 Eppelheim Germany 49 6221 765767 49 6221 764620 fax info polysciences de Asia Pa...

Reviews: