'Settings'
allows the user to choose the location of
the saved reports on the PC. Dragging and dropping
the yellow boxes containing 'Day', 'Month' etc. into the
empty boxes above will create a sub structure within the
save directory. Up to 3 levels of sub structure can be
defined in this way.
If for instance you select 'Day' (as seen left) a new sub
folder will be created whenever a report is saved on a
new day. If 'Checklist' or 'Saf-T-Log device' is used the
reports will be filed according to matching checklists or
Saf-T-Log devices.
'Settings' also allows the user to opt in or out of purging
the readings from the Saf-T-Log during the automatic
synchronising.
Defaults for UK or US time format, °C or °F temperature
units and time zone can also be adjusted here for new
checklists that may be created. Click 'Save Settings' to
keep the changes. Click 'Cancel' to discard changes.
Click 'Configure Saf-T-Log' to assign a name or ID tag
to a Saf-T-Log thermometer or adjust ('Sensor Trim')
the Saf-T-Log. The serial number of the Saf-T-Log
thermometer that is connected is displayed at the top of
this window. Once the desired settings are completed,
click 'Update' to immediately update the connected
Saf-T-Log thermometer.
NOTE: It is NOT recommended to adjust
the 'Sensor Trim' value without appropriate
temperature standards and experience. Call
ETI Technical Sales on 01903 202151 for more
details.
The 'Configure Saf-T-Log device' window can be
password protected to lock users from accessing this
window. A password can be created by clicking the
'Password Protection' button.
SETTINGS
CONFIGURE SAF-T-LOG
8