User manual
PONSEL
Référence : NOTICE_ACTEON_2050_v001
V : 002
21/67
Press the ENTER key to confirm the first
calibration point
CALIBRATION
OF SS SENSOR
Immerse the sensor in
CLEAN WATER
SS content:
Measure:
g/l
g/l
ESC: Cancel the procedure
ENTER: Validate the zero
0.00
0.15
WAIT: Nonstable measurement
AWAIT MEASUREMENT STABILISATION
Use the
and
keys to adjust
sludge concentration
Press the ENTER key to confirm the second
calibration point
AWAIT MEASUREMENT STABILISATION
Take a representative sample of the measurement medium,
enough to immerse the sensor (in a 5 litre pail or larger).
Dip the sensor in the half-filled pail of sludge and keep
stirring gently to prevent decanting and to keep the SS in
suspension as much as possible.
Advice: The sample sludge must remain at the temperature
of the tank and must not be active
calibrate immediately
after sample taking.
Calibration
Correct
If the first calibration point is incorrect, an error
message window appears (see "Error message
information" chapter §6.2.1)
No
Yes
CALIBRATION
OF SS SENSOR
Immerse the sensor in
your SLUDGE
g/l
g/l
ESC: Leave the current slope
ENTER: Validate the new slope
5.00
4.20
WAIT: Nonstable measurement
Current slope coefficient: 100 %
New slope coefficient: 92 %
SS content:
Measure: