CALIBRATION
SENSORS CALIBRATION
SS SENSOR
Gain adjust
Immediate calibration
Differed calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to select
the type of calibration and the
probe to be calibrated.
Select the
Immediate calibration
menu then
press ENTER
.
Use the
and
keys to move the
cursor in the name section.
Use the keys
and
to change the
letters. (The scrolling order of the
letters is A…Z,0..9,?,>,space )
Enter your name and reference and press ENTER
CALIBRATION
SS SENSOR
ESC: Cancel the procedure
ENTER: Start calibration
Enter your name:
D
urant
Dip the sensor in clean water with no bubbles (tap water
for example). Preferably, this water must be at the same
temperature as the tank sludge in which the probe will be
placed.
Stir the water with the probe to disperse any bubbles on
the optic windows.