User manual
PONSEL
Référence : NOTICE_ACTEON_2050_v001
V : 002
44/67
The following message will be displayed if the offset is incorrect:
Press the ENTER key then restart SS sensor offset
adjustment.
CONFIG
WHAT DO YOU WANT TO DO?
Averaging sensor response
Sludge slope coefficient reminder
Sensor offset adjustment
Graphic screen
4-20mA outputs adjustment
Relay adjustment
Measure units
Language
Factory default value reminder
ESC: Escape menu
CONFIG
Cancellation of offset impossible!!
Check the sensor (the ruler might not
be completely opaque and well
positioned in the optical slot, between
the two port-holes of emission and
reception of the infra-red beam)
ENTER: Escape menu
EXTINCTION OF THE I.R.LIGHT
BEAM:
!!! OUT OF RANGE!!!
Comment: The SS sensor
offset has not been updated.
Your device is using the last
correct offset.