User manual
PONSEL
Reference: NOTICE_ACTEON_2010_v001-UK
V: 001
10/56
•
connection sensor – transmitter > 10 m
Comment:
If the connection cable between the pH sensor and transmitter is longer than 10 metres, a leak-tight IP 68
connection box with integrated pre-amplifier must be used (REF: PON-BJAI-E PH/EH/°C ).
Figure 3– Installation including a connection box with integrated pre-amplifier
2.2.1 Acteon 2010 wiring
See appendix (§15) at the end of the document.
2.3 Tank-mounting
2.3.1 Using the stand and protective hood
A protective hood (PON-PDPVC-1) is available for mounting the ACTEON 2010.
The hood is essential in the case of direct exposure to adverse weather or sunshine.
Figure 4– Mount and hood for the transmitter
ACTEON 2010 pH and Temperature
transmitter
Temperature sensor
PONCPC-T-10
230 V~
Relay
4-20 mA
Max. 10 metres
Connection box
PON-BJAI-E PH/EH/°C
pH sensor
PONCPC-PH-10
Earth terminal
or lining-out of
230 V