User manual
PONSEL
Reference: NOTICE_ACTEON_2010_v001-UK
V: 001
21/56
Use the
and
keys to adjust
the value of the buffer.
To validate the first calibration point, press the
ENTER key.
CALIBRATION
OF THE SENSOR pH
Immerse sensor in
a 7.01 pH buffer
Standard:
Measure:
:
pH
pH
ESC: Cancel the procedure
ENTER: Validate pH 7
7.01
7.15
WAIT, Nonstable measurement
MEASUREMENT STABILISATION.
Use the
and
keys to adjust
the value of the buffer (from
4.98 to 10.13).
To validate the second calibration point, press the
ENTER key.
CALIBRATION
OF THE SENSOR pH
Immerse sensor in
a pH 4.01 or 10.01 buffer
Standard
:
Measure
pH
pH
ESC: Leave the current gain
ENTER: Validate the new gain
4.01
4.20
WAIT, Nonstable measurement
MEASUREMENT STABILISATION .
Immerse the sensor in the pH 4.01 or 10.01 buffer solution.
Calibration
Correct
If the first calibration point is not correct, an error
message window appears (see "information on error
messages” chapter §6.2).
No
Yes