6.1.4 Complete calibration of the temperature sensor (at two points)
The following procedure enables complete calibration of your temperature sensor.
A precision thermometer and water at 0°C are necessary for this procedure.
CALIBRATION
SENSORS CALIBRATION
SENSOR pH
Gain adjust
Complete calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to
select the type of calibration
and sensor to be calibrated.
Select the
Complete calibration
menu for the
temperature sensor then press the ENTER key.
Use the
and
keys to move in
the name.
Use the
and
keys to change the
letters.
Once you have entered your name or reference, press
the ENTER key.
CALIBRATION
OF TEMPERATURE SENSOR
ESC: Cancel the procedure
ENTER: Start calibration
Enter your name:
D
urant
Immerse the sensor in a 0.00°C solution of water
and crushed ice.