To validate the first calibration point, press the
ENTER key.
CALIBRATION
OF TEMPERATURE SENSOR
Immerse sensor in
water + ice
Standard:
Measure:
:
°C
°C
ESC: Cancel the procedure
ENTER: Validate the zero
0.00
0.15
WAIT, Nonstable measurement
MEASURELENT STABILISATION.
Use the
and
keys to adjust the
value of the water temperature.
To validate the second calibration point, press the
ENTER key.
CALIBRATION
OF TEMPERATURE SENSOR
Immerse sensor in
water at ambient temperature
Standard:
Measure:
:
°C
°C
ESC: Leave the current gain
ENTER: Validate the new gain
20.00
20.20
WAIT, Nonstable measurement
AWAIT MEASUREMENT STABILISATION.
Immerse the sensor in the water at ambient temperature.
Calibration
Correct
If the first calibration point is not correct, an error
message window appears (see "information on error
messages” chapter §6.3.1).
No
Yes