User manual
PONSEL
Reference: NOTICE_ACTEON_2010_v001-UK
V: 001
29/56
6.1.5 Adjusting the slope of the temperature sensor
If you observe a small measurement error, you may only adjust the slope of your sensor by implementing the
following procedure:
CALIBRATION
SENSORS CALIBRATION
SENSOR pH
Gain adjust
Complete calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to
select the type of calibration
and sensor to be calibrated.
Select the
Gain adjust
menu for the temperature sensor then
press the ENTER key.
Use the
and
keys to adjust
the temperature value of the
water at ambient temperature.
To validate the gain adjustment, press the ENTER
key.
CALIBRATION
OF TEMPERATURE SENSOR
Immerse sensor in
water at ambient temperature
Standard:
Measure:
°C
°C
20.00
20.20
WAIT, Nonstable measurement
MEASUREMENT STABILISATION.
Immerse the sensor in the water at ambient temperature.
ESC: Leave the current gain
ENTER: Validate the new gain