6.2 Information on error messages during calibration of the pH sensor
6.2.1 Error during calibration with the pH 7.01 buffer
If you observe an error during calibration with the pH 7.01 buffer:
1)
Check that the buffer is not fouled or out-of-date.
2)
Check that the pH bulb of the sensor is not soiled.
6.2.2 Error during calibration with the pH 4.01 buffer
If you observe an error during calibration with the pH 4.01 buffer:
1)
Check that the buffer is not fouled or out-of-date.
2)
Check that the port for the bridge is not blocked.
6.3 Information on error messages during calibration of the temperature sensor
6.3.1 Error during calibration at 0°C.
CALIBRATION
OF THE TEMPERATURE SENSOR
CALIBRATION IMPOSSIBLE
offset:
NON CORRECT
Esc: Escape the menu
CALIBRATION
Esc: Escape the menu
OF SENSOR pH
CALIBRATION IMPOSSIBLE
Slope:
NON CORRECT
CALIBRATION
OF SENSOR pH
CALIBRATION IMPOSSIBLE
Offset:
NON CORRECT
Esc: Escape the menu