8 VIEWING THE SENSOR CALIBRATION REPORT
The calibration logger enables the reports for the most recent calibrations to be checked.
To access this information follow the instructions below:
This window allows a diagnostic of whether any errors have occurred in the most recent calibrations.
List of diagnostic messages from the last calibration:
Message
Explanation
CORRECT
ACTEON 2010 has been calibrated correctly.
THEORETICAL coef.
ACTEON 2010 is using the theoretical coefficients (slope and offset); this
message is displayed after a theoretical calibration.
THEORETICAL slope
ACTEON 2010 is using the theoretical slope.
THEORETICAL offset
ACTEON 2010 is using the theoretical offset.
INCORRECT slope
ACTEON 2010 is incorrectly calibrated because the slope calculated during
calibration is incorrect (in this case ACTEON 2010 is using the last correctly
calculated slope).
INCORRECT offset
ACTEON 2010 is incorrectly calibrated because the offset calculated during
calibration is incorrect (in this case ACTEON 2010 is using the last correctly
calculated offset).
20.00
°C
-24h
4.31
MEASURE
pH
INFO
CONF
HIST CAL
LOG
MEAS
ACTEON 2010 pH meter
pH
6
8
Use the
and
keys to
move about the ACTEON
menu.
Select the
menu then press the ENTER key.
LOG
CALIB LOG
CALIBRATION LOGGER
Sensor Calibration
pH
CORRECT
Standard 1: 4.01pH
Standard 2: 7.01pH
Offset: 0.03pH
Slope: 110%
Name: Ponsel
Sensor Calibration
Temperature
CORRECT
Standard 1: 0.01°C
Standard 2: 20.00°C
Offset: 1.65°C
Slope: 110%
Name: Ponsel
Esc: Escape the menu
Mode1 R1
R2