9.3.1 Adjusting the stopped thresholds for the 4-20mA outputs
Use the
and
keys to select
the desired configuration.
Select the
4-20mA output adjustment
menu then press
the ENTER key.
Use the
and
keys to increase and
decrease the values of the 4 and 20mA
stops.
Use the
and
keys to select either
the 4mA or 20mA stop.
CONFIG
4-20mA OUTPUTS
4-20mA output adjustment
4-20mA output calibration
CONFIG
4-20mA OUTPUT
NUMBER 1 FOR pH
4mA adjustment:
2.00
pH
20mA adjustment:
12.00 pH
Once the 4-20mA output is correctly configured for the
pH, press the ENTER key.
Use the
and
keys to increase
and decrease the values of the 4 and
20mA stops.
Use the
and
keys to select
either the 4mA or 20mA stop.
CONFIG
4-20mA OUTPUTS
NUM 2 FOR TEMPERATURE
Esc: Escape the menu
Enter: Validate the choice
Esc: Escape the menu
Enter: Validate the choice
4mA adjustment:
-5.00
°C
20mA adjustment:
35.00 °C