9.4 Adjusting the relay outputs
The 2 relays can be configured in 2 different modes :
- Mode 1 enables configuration of relay R1 on a pH measurement value and relay R2 on a temperature
measurement value.
- Mode 2 enables configuration of relays R1 and R2 on two pH measurement values.
The 2 relays can be configured with the following attributes:
-
Alarm threshold: threshold trigger value
-
Hysteresis: hysteresis value for toggling of relays (the hysteresis enables pumping phenomena
on the relays to be avoided)
-
Triggering direction:
Above: this means the relay’s contact is closed above the alarm threshold
Below: this means the relay’s contact is open below the alarm threshold
Use the
and
keys to select
the desired configuration.
Select the
Relays outputs adjustment
menu then press
the ENTER key.
Use the
and
keys to select the
desired configuration mode.
CONFIG
CONFIG
WHAT DO YOU WANT TO DO?
Averaging sensor response
Graphic screen
4-20mA output adjustment
Relays outputs adjustment
Measure unit
Language
Factory default value reminder
CONFIGURATION OF THE RELAYS
MODE:
- mode 1: alarms adjustment
- mode 2: Regulation 2 Relays