Subsequent to a factory default value reminder the ACTEON 2010 will be configured with the following
parameters:
Averaging sensor response
10 measures
Time base for display of the trend
curve
1 minute
4-20mA outputs adjustment
pH output: The 4mA stop is set to 2 pH and the 20mA stop is set to 12 pH
Temp. °C output: The 4mA stop is set to -5°C and the 20mA stop is set to
35°C.
Relay outputs adjustment
The relays are configured in mode 1:
R1: threshold: pH 7.00; hyst: 0.20 above the threshold, the R1 relay is
active.
R2: threshold: 50.00°C; hyst: 0.10 above the threshold, the R2 relay is
active.
Measurement unit for temperature
°C
Language
French
10 The info menu
The INFO window provides information on the type and version of the device. The software and hardware
versions must be mentioned when calling technical support or the Ponsel after sales service.
11 ADJUSTING THE CONTRAST OF THE ACTEON 2010 DISPLAY
The contrast of the LCD screen can be adjusted in order to modify the display. This adjustment can only be
carried out from the measure window.
INFO
ACTEON 2010 pHmeter
Temperature
Software version: XX.XX
Hardware version: XX.XX
Esc: Escape the menu
20.00
°C
-24h
4.31
MESURE
pH
INFO
CONF
HIST CAL
LOG
MEAS
ACTEON 2010 pH meter
pH
6
8
Press the ESC and
keys
simultaneously to reduce the
contrast.
Press the ESC and
keys
simultaneously to increase the
contrast.
Mode1 R1
R2