14 Questions & Answers
14.1 pH measurement anomalies
The response to pH variations is too weak:
The glass electrode is soiled and must be cleaned (rub carefully with filter paper soaked with toluene).
There is no response to pH variations:
The glass electrode bulb is broken or cracked, or the connection between the sensor and ACTEON 2010
is faulty.
The measurement is unstable:
The electrolyte contact with the medium is poor, the connection between the sensor and ACTEON 2010
is faulty or the connector is wet: rinse thoroughly with alcohol.