Functionality
Used to access the sensor’s calibration menu.
This menu is used to apply the default calibration coefficients set in the factory.
Detailed calibration log: this menu contains the data for the last ten calibrations
(offset/gain) … The coefficients for a specific calibration can be reused.
The
Calibration settings
window lets the operator: select the parameter to be calibrated; enter the
operator's name; select a calibration mode and set the values of the standards used.
Set-up line
Functionality
Parameter
The parameter to be calibrated can be selected from a drop-down menu which
offers the parameters measured by the sensor.
Operator
This menu is used to enter the Operator's name (maximum of 13 characters)
which will be saved with the results of a calibration process.
Type
This line appears if the parameter selected for calibration is oxygen as a %Sat.
To complete this line, the operator can select either a 2-point calibration
(OFFSET and GAIN) or a single-point calibration (GAIN only).
Mode
The options are: "Single" or "Multiple" mode: when several sensors of the same
type are connected, a parameter common to all the sensors can be calibrated.
The "Single" option is used when calibrating one parameter for one sensor, and
"Multiple" is used when calibrating one parameter for 2 sensors (e.g. the
Temperature parameter measured by all the sensors, or the connection of 2 pH
sensors).
Standard 1
In the "Standard 1" window, which only appears if the value can be modified, the
value of the Standard can be set using a number pad and will be used during the
first step of the calibration for the calculation of the OFFSET.
The value which can be set cannot exceed the minimum and maximum values.