Addressing of the digital sensors.
The first 3 lines of the "Addressing" window are used to define the serial number of the sensor affected
by the change of address.
After configuring the Serial No. and address, confirm with the
icon.
The sensor's serial number is engraved on the body of the sensor and comprises the following
information:
Set-up line
Functionality
Type SN-XXXX
PODO for an OPTOD sensor
PNEP for a Turbidity sensor
PC4E for a C4E sensor
PPHR for a PHEHT sensor
PORP for an ORP sensor
PTUR for a VB5
– MES 5 sensor
Revision
Represented by a letter (A in the example above)
Number
A 4-digit number (0000 in the example above)
Address
Between 1 and 243
After configuring the Serial No. and address, confirm with the
icon. The change made to the
Modbus address becomes effective when the "Communication completed successfully" message
appears.
To access the "Addressing" menu, select the
icon in the Sensors scan window or select the
icon in the Numeric input settings screen.