36
Version 2.1
Update: April 2016
PONSEL Mesure
– Aqualabo Contrôle
35 Rue Michel Marion
– 56850 CAUDAN, FRANCE
Tel.: +33(0)2 97 89 25 30
– Fax: +33(0)2 97 76 55 72
Lines 3 to 32
: a number of conversion points in the form of portions of a
straight line Source signal 2 = f(Source signal 1).
To activate the conversion, plug the stick into the ACTEON 5000's USB port
(refer to item 6 of diagram 3) and wait for the small green LED located
above the USB port to stop flashing.
Then navigate to menu I2: Source signal 2 and select the "Active" option.
If Source signal 2 is Activated
I2: Source signal 2
display
Not displayed
Sens 1. Prim. Meas: the measurement shall be displayed in the space
allocated to the primary parameter for sensor 1 (item 3 on the home screen)
Sens 1. Sec. Meas: item 9 on the home screen
Sens 2. Prim. Meas: item 6 on the home screen
Sens 2. Sec. Meas: item 13 on the home screen.
I2: Source signal 2
stability delta
Can be set from 0 to 50 % (the default is 1 %) and is used to define when
the arrow on the home screen will indicate that the measurement is stable if
the parameter measured has been set as the primary parameter for sensor
1 or 2 (refer to items 2 and 5 in Table 2)