Power Probe III S Speci
fi
cations
Circuit Breaker
8 amp thermal response - Manual reset
Typical Response
8 amp 10 amps 15 amps 25 amps Short Circuit
No Trip 20 min. 6 sec. 2 sec. 0.3 sec.
13
Short
Circuit
25 Amps
15 Amps
10 Amps
8 Amps
0.1 0.3 2 sec. 6 sec. 20 min.
DC Voltage ............... 0 to +70 Volts +1 digit
P-P Voltage .............. 0 to +70 Volts
Frequency Response ......... 10Hz to 10kHz
(for tone pass through)
P-P display ........... 15Hz Square Wave
35Hz Sine Wave
DC Voltmeter Mode - Continuity to ground
- First Level - display is enabled less than 20K
Ω
- Second Level - green LED is enabled less than
approx. 650
Ω
- & + Peak Detector Response
- Single event capture less than 200mS pulse width
- Repetitive events less than 1mS pulse width
Peak to Peak Mode ....... 0 to +70 Volts + 1 digit
- 4Hz to over 500kHz Square Wave input
Over Volt Warning
If the probe tip connects to a voltage
greater than +70 Volts the display
will show an “Over Volt” warning.
Remove the probe tip immediately to
prevent internal damage to the tool.
Over Load Warning
If the probe is connected to a bat-
tery with voltage greater than 34VDC
the display will show an “Over Load”
warning.
Storage temperature/humidity: -20 to 70
℃
,
70% RH max
Operating temperature/humidity: -10 to 50
℃
,
70% RH max
Pollution degree: 2