Med Touch Series LCD Alarm Systems
Powerex
•
150 Production Drive
•
Harrison, OH 45030
•
USA
P 1.888.769.7979
•
F 513.367.3125
•
www.powerexinc.com
IN597400AV
•
08/2016
Page 12 of 56
Wiring the Remote Devices to Remote Signal Master Board
(4-20mA Transmitters)
Locate 3 port terminal connector on the right edge of the back of
the remote signal (Master) board.
The terminal connector will be labeled 4-20mA.
The terminal connector may be un-plugged from the circuit board
to simplify installation of the wires.
For 4-20mA transmitters with loop power supplied by an external
source.
Land the 4-20mA signal wire in the Number 2 terminal port. Land
the (-) ground wire to the Number 3 terminal port.
For 4-20mA transmitters with loop power supplied by the Alarm
Circuit Board (35-0037)
Land the 4-20mA signal wire to the Number 2 terminal port. Land
the +12VDC wire to the Number 1 terminal port.
Wiring General Fault Dry Contacts to Building Automation
System or Remote Alarm
The button board (shown here) is located in the top left corner of
the alarm panel. The dry contacts which provide the general fault
alarm condition are located on the bottom edge of the board.
Normally Open (NO), Normally Closed (NC) and Common (C)
connections are provided.
The terminal connector may be un-plugged from the circuit board
to simplify installation of the remote signal wires.