P2
POWERTRONIX Corporation
Page
4
of
10
WARNING:
In an emergency, main electric supply to the unit may cut off by unplugging the power
cord A] Either from the wall outlet or from the appliance coupler of the units supplied
with an appliance coupler and B] from the wall outlet only for the units supplied with a
non-detachable power cord. Some models are offered with an on/off switch. If you
are using one of these unites, turn the power switch OFF.
WARNING:
Do not expose to temperatures below 0° Celsius (32°F) or above 65° Celsius (149°F).
Do not expose this apparatus to relative humidity more than 95% RH (non-
condensing).
Do not expose to pressure levels below 68.6 kPa or above 101.3 kPa.
WARNING:
This apparatus is a Class 1 equipment.
WARNING:
Do not modify Isolation Station unit at any time during its entire service life. If the
Isolation Station unit is modified by any manner for any purpose during its service life,
user needs to carry out an evaluation to the requirements of the IEC 60601-1
standard and relevant safety standards for the end application.
WARNING:
This unit is a multiple socket outlet [MSO] device. When user connects load
equipment to Isolation Station (MSO) the total system becomes a new ME System
and user needs to carry out an evaluation to the requirements of IEC 60601-1
standard for the total system.
POWER CORD REQUIREMENTS
Input
: All power cords must be appropriately approved by testing agencies in the country where the
Isolation Station
is Used, such as UL/CSA/VDE/SEMKO, etc. For the US and Canadian market, a
HOSPITAL GRADE plug is required for medical applications.
One end (plug) must be appropriate for the wall receptacle, while the other end is made to fit into the
power inlet of the
Isolation Station
, this end must be of appliance type, IEC-320 female plug. All power
cords must have grounding.
POWER
AWG (NOMINAL CROSS-
SECTIONAL AREA mm
2
Cu)
RECOMMENDED LENGTH
300VA
AWG 18 (0.75mm
2
)
Less Than 12 Feet
600VA
AWG 18 (0.75mm
2
)
Less Than 12 Feet
800VA
AWG 16 (1.25mm
2
)
Less Than 12 Feet
1000VA
AWG 16 (1.25mm
2
)
Less Than 12 Feet
1150VA
AWG 14 (2.0mm
2
)
Less Than 12 Feet
1500VA
AWG 14 (2.0mm
2
)
Less Than 12 Feet
1800VA
AWG 14 (2.0mm
2
)
Less Than 12 Feet
2400VA
AWG 12 (3.0mm
2
)
Less Than 12 Feet
3000VA
AWG 10 (5.0mm
2
)
Less Than 12 Feet