4104V102-UK 11
Input specifications
Current input
Signal range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±23 mA
Programmable measurement ranges . . . . . . . . . . . . . . . . . . . . . . . . 0...20, 4...20, ±10, ±20 mA
Input voltage drop, nom. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4 V @ 20 mA
Loop error detection, 4...20 mA:
Low detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < 3.6 mA
High detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . > 21 mA
Voltage input
Signal range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±12 VDC
Programmable measurement ranges . . . . . . . . . . . . . . . . . . . . . . . . 0...1, 0.2...1, 0...5, 1...5, 0...10, 2...10 V,
±1, ±5 and ±10V
Input resistance, nom.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 MΩ
Current output specifications
Active unipolar and bipolar mA
Programmable ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-20, 4-20, ±10 and ±20 mA
Direct or Inverted action
V-curve function, 100-0-100% . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-0-20 mA
Load, max.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 800 Ω
Passive 2-wire mA
Programmable ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0...20 and 4...20 mA
Direct or Inverted action
V-curve function, 100-0-100% . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-0-20 mA
External 2-wire loop supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5...26 V
Common specifications current output
Signal range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0...23 mA (unipolar) / -23...+23 mA (bipolar)
Current limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ≤ 28 mA (unipolar) / ±28 mA (bipolar)
Load stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ≤ 0.001% of span / 100 Ω
Response time, programmable. . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.0...60.0 s
Output limitation at outside range:
on 4...20 and 20...4 mA signals . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8...20.5 mA
on other unipolar signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 and 115% of max. value
on bipolar signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±115% of min. and max. values
Sensor error indication, at 4...20 mA input:
selectable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Low, High, Zero or None
Low - corresponds to 0 mA at 0...20 mA and to 3.5 mA at 4...20 mA
High - corresponds to 23 mA at both 0...20 and 4...20 mA
Zero - equals 0 mA output
None - the output state is undefined
Voltage output specifications
Programmable ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0/0.2...1, 0/1...5, 0/2...10, ±1, ±5 and ±10 V
Direct or Inverted action
V-curve function, 100-0-100% . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-0-1 V, 5-0-5 V and 10-0-10 V
Load, min. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . > 500 kΩ
Response time, programmable. . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.0...60.0 s
Output limitation at outside range:
on unipolar signals starting from 0 . . . . . . . . . . . . . . . . . . . . . . . . 0 and 115% of max. value
on unipolar signals with offset . . . . . . . . . . . . . . . . . . . . . . . . . . -5% of min. value and 115% of max. value
on bipolar signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±115% of min. and max. values
Sensor error indication, at 4...20 mA input:
selectable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Low, High, Zero or None
Low - corresponds to the selected min range value
High - corresponds to the selected max range value
Zero - equals 0 V output
None - the output state is undefined