9106 - Product version 9106-002
13
Visualisation in the PR 4500xx of hardware / software error
Readout at hardware error
Error search
Readout
Cause
Communications test PR 4500 and 9106
NO.CO
Connection error
EEprom error - check configuration
FL.ER
Configuration error or CRC mismatch, recovery
configuration is loaded
User error
II ! / II !
Loop limit exceeded
User error
II ! / II !
Loop error
EEprom error - check configuration
EE.ER / IE.ER Invalid configuration (CRC or data)
Hardware error
SU.ER
Supply error
Hardware error
RA.ER
RAM error
Hardware error
FL.ER
Flash error
Hardware error
IN.ER
Initialization error
Hardware error
C1.ER
Hardware error - channel 1
Hardware error
C2.ER
Hardware error - channel 2
Hardware error
DE.ER
General error
! All error indications in the display flash once per second. The help text explains the error.
In case of cable fault the backlight also flashes. This can be reset by pressing the
3
key.
Errors affecting both channels are shown as error on channel 1 - and the line showing channel 2 is blank.
Hardware error can be reset in two ways. Either step through the menus (if the other channel is to stay in
operation) or power cycle the device.