9106QI01
LERBAKKEN 10, 8410 RØNDE DENMARK
Revision date:
2020-06-18
Version Revision
V6 R0
Prepared by:
PB
Page:
2/6
Hazardous area Non Hazardous area
Zone 0,1,2, 20, 21, 22
or Zone 2
9106B1A, 9106B1B
Ex input : Loop current source
CH1 (terminal 43,44)
CH2 (terminal 53,54)
U
o
:
27.5 V
I
o
:
92.6 mA
P
o
:
0.64 W
IIC
IIB
IIA
I
C
o.
0.084
F 0.670
F
2.23
F
3.94
F
L
o.
4.15 mH 16.59 mH
33.17 mH
54.42 mH
L
o/
R
o
--------
223 µH/Ω 447 µH/Ω
733 µH/Ω
-20 ≤Ta ≤ +60ºC
(terminal 11,12,13,14)
(terminal 31,32,33,34)
(terminal 91,92,93,94,95)
U
m
: 253V, max 400Hz
44
43
42
41
54
53
52
51
34
33
32
31
14
13
12
11
9106
4501
91 92 93 94 95
Power
Rail
CH1
CH2
T
T
_
+
+
_
9106B2A, 9106B2B
Ex input : Loop current source
CH1 (terminal 43,44)
CH2 (terminal 53,54)
U
o
:
25.3 V
I
o
:
96 mA
P
o
:
0.61 W
IIC
IIB
IIA
I
C
o.
0.104
F 0.818
F 2.85
F
4.74 µF
L
o.
3.86 mH
15.43 mH 30.86 mH
50.64 mH
L
o
/ R
o
---------
234 µH/Ω 468 µH/Ω
769 µH/Ω
22
9106 - Product version 9106-002