10
5114V107-UK
Electrical specifications, temperature input:
TC input:
Max. offset ................................................................. 50% of selec. max. value
Sensor error current ................................................ Nom. 30 µA
CJC ..................................................................................
≤
±1°C
Sensor error detection ........................................... Yes
mV input:
Measurement range ................................................ -150...+150 mV
Min. measurement range ...................................... 5 mV
Max. offset ................................................................. 50% of selec. max. value
Input resistance........................................................ Nom. 10 M
Ω
RTD and linear resistance input:
Max. offset ................................................................. 50% of selec. max. value
Max. cable resistance per wire ........................... 10
Ω
Sensor current ........................................................... Nom. 0.2 mA
Effect of sensor cable resistance
(3- / 4-wire) ................................................................ < 0.002
Ω
/
Ω
Sensor error detection ........................................... Yes
Type
Min. value
Max. value
Min. span
Standard
Pt100
Ni100
Lin. R
-200°C
-60°C
0
Ω
+850°C
+250°C
5000
Ω
25°C
25°C
30
Ω
IEC 751
DIN 43760
-------
Type
Min.
temperature
Max.
temperature
Min.
span
Standard
B
E
J
K
L
N
R
S
T
U
W3
W5
LR
+400°C
-100°C
-100°C
-180°C
-100°C
-180°C
-50°C
-50°C
-200°C
-200°C
0°C
0°C
-200°C
+1820°C
+1000°C
+1200°C
+1372°C
+900°C
+1300°C
+1760°C
+1760°C
+400°C
+600°C
+2300°C
+2300°C
+800°C
200°C
50°C
50°C
50°C
50°C
100°C
200°C
200°C
50°C
75°C
200°C
200°C
50°C
IEC584
IEC584
IEC584
IEC584
DIN 43710
IEC584
IEC584
IEC584
IEC584
DIN 43710
ASTM E988-90
ASTM E988-90
GOST 3044-84