background image

P

RO

V

U

 PD6210 Analog Input Batch Controller Instruction Manual

 

 

 

 

 

 

42 

 

Advanced Features Menu & Display Messages 

The following table shows the functions and messages of the 

Advanced Features

 menu in the order they 

appear in the menu. 

Display 

Parameter  Action/Setting 

Lo G 

Low gate 

Program Low gate value 

Hi G 

High gate 

Program High gate value 

filter 

Filter 

Set noise filter value 

 Hi spd 

High Speed 

Filter 

Program High Speed 

Filter 

 Lo spd 

Low Speed 

Filter 

Program Low Speed 

Filter 

bypASs 

Bypass 

Set filter bypass value 

Round 

Round 

Set the rounding value 

for display variables 

seriAl 

Serial 

Set serial 

communication 

parameters  

SlaveId 

Slave ID 

Set slave ID or controller 

address  

baud 

Baud rate 

Select baud rate 

Tr dLY 

Transmit  

delay 

Set transmit delay for 

serial communication 

Parity 

Parity 

Select parity 

Even, Odd, or None with 

1 or 2 stop bits 

t-byt 

Time byte 

Set byte-to-byte timeout 

SelEct 

Select 

Enter the Select menu 

(function, cutoff, out) 

Functn 

Signal Input 

Conditionin

Select linear, square 

root, or programmable 

exponent 

 Linear 

Linear 

Set controller for linear 

function and select 

number of linearization 

points 

  No PtS 

Number of 

points 

Set controller for 2 to 32-

point linearization 

 Square 

Square root 

Set controller for square 

root extraction 

 Prog E 

Programma

ble 

exponent 

Set controller for 

programmable exponent 

and enter exponent 

value 

CutofF 

Cutoff 

Set low-flow cutoff 

Count 

Count 

Set the totals to count up 

or down from a set value 

 Tot C 

Total count 

Set the batch to count up 

to or down from the 

preset 

  up 

Up 

Set the total to count up 

Display 

Parameter  Action/Setting 

  Douwn 

Down 

Set the total to count 

down 

 Gtot C 

Grand total 

count 

Set the grand total to 

count up or down from a 

defined value 

  C strt 

Count start 

Set the starting grand 

total value when 

counting down 

AoutPr 

Analog 

output 

programmi

ng 

Program analog output 

parameters 

 Source 

Source 

Select source for the 4-

20 mA output 

 O-rang 

Overrange 

Program mA output for 

display overrange 

 u-rang 

Underrange 

Program mA output for 

display underrange 

 Break 

Loop Break 

Set relay condition if loop 

break detected 

  Force 

Force 

output 

Force analog output 

value for loop break 

 Ignore 

Ignore 

Ignore loop break 

condition 

 nmAx 

Maximum 

Program maximum mA 

output allowed 

 nmin 

Minimum 

Program minimum mA 

output allowed 

 Calib 

Calibrate 

Calibrate 4-20 mA 

output (internal 

reference source used 

for scaling the output) 

  4 nmA 

4 mA 

output 

Enter mA output value 

read by milliamp meter 

with at least 0.001 mA 

resolution 

  20 nmA 

20 mA 

output 

Enter mA output value 

read by milliamp meter 

with at least 0.001 mA 

resolution 

user 

User I/O 

Assign function keys and 

digital I/O 

F1 

F1 function 

key 

Assign F1 function key 

F2 

F2 function 

key 

Assign F2 function key 

F3 

F3 function 

key 

Assign F3 function key 

Summary of Contents for PD6210-6H2

Page 1: ...age Batching with up to 8 Relays Automatic Overflow Protection Manual Control or Automatic Batching Dual Line Display NEMA 4X and IP65 Rated Front Panel UL Listed CE Marked Display Features 0 6 0 46 D...

Page 2: ...s Anyone using this product for such applications does so at his her own risk Precision Digital Corporation shall not be held liable for damages resulting from such improper use Limited Warranty Preci...

Page 3: ...Setting the Relay Operation relay 31 Relay Assignment Assign 32 Setting the Relay Action 32 Setting Batch Control Relays 32 Programming Alarm Set and Reset Points 33 Setting Fail Safe Operation 33 Pro...

Page 4: ...uperluminous LED digits make it easily readable in smoke dust fog and with the optional SunBright display even direct sunlight The controller comes programmed for easy front panel start pause stop and...

Page 5: ...210 6H4 PD6210 7H4 4 relays PD1104 PD6210 6H5 PD6210 7H5 2 relays 4 20 mA output PD1105 PD6210 6H7 PD6210 7H7 4 relays 4 20 mA output PD1107 Model number for replacement option card Accessories Model...

Page 6: ...al Password Prevents resetting the total manually Gtotal Password Prevents resetting the grand total manually Non Volatile Memory All programmed settings are stored in non volatile memory for a minimu...

Page 7: ...99 9 seconds applied to the first relay assigned to total or grand total If the controller is programmed to reset total to zero automatically when the preset is reached then a delay will occur before...

Page 8: ...wer the 4 20 mA output or other devices Refer to Figure 6 on page 12 and Figure 14 on page 15 All models rated 40 mA max External Loop Power Supply 35 VDC maximum Output Loop Resistance Power supply M...

Page 9: ...munity standard RFI Amplitude Modulated 80 1000 MHz 10 V m 80 AM 1 kHz 1 4 2 0 GHz 3 V m 80 AM 1 kHz 2 0 2 7 GHz 1 V m 80 AM 1 kHz Electrical Fast Transients 2kV AC mains 1kV other Electrostatic Disch...

Page 10: ...ry for assistance Panel Mounting Instructions Prepare a standard 1 8 DIN panel cutout 3 622 x 1 772 92 mm x 45 mm Refer to Figure 1 for more details Clearance allow at least 6 0 152 mm behind the pane...

Page 11: ...e factory configured to provide 24 VDC power for the transmitter or sensor If the transmitter requires 5 or 10 VDC excitation the internal jumper J4 must be configured accordingly To access the voltag...

Page 12: ...e connectors label affixed to the controller shows the location of all connectors available with requested configuration Warning Do not connect any equipment other than Precision Digital s expansion m...

Page 13: ...voltage inputs Setup and programming is performed through the front panel buttons Figure 7 Transmitters Powered by Internal Supply Figure 8 Transmitter Powered by Ext Supply or Self Powered The curren...

Page 14: ...loads to prevent disrupting the microprocessor s operation The suppressors also prolong the life of the relay contacts Suppression can be obtained with resistor capacitor RC networks assembled by the...

Page 15: ...signal applied to F4 Figure 13 F4 Digital Input Connections 4 20 mA Output Connections Connections for the 4 20 mA transmitter output are made to the connector terminals labeled MA OUT The 4 20 mA out...

Page 16: ...ules DIN Rail Mounting Kit Figure 16 External Relays Module Connections Figure 17 Digital I O Module Connections Interlock Relay Feature As the name implies the interlock relay feature reassigns one o...

Page 17: ...t Panel Buttons and Status LED Indicators Button Symbol Description LED Status Menu 1 8 Relay 1 8 indicators START Right arrow F1 R Rate indicator BATCH Up arrow F2 T Batch Total indicator STOP Enter...

Page 18: ...r as alarms that will be active when the batch process is running A 4 20 mA output option is also available for retransmitting the process variable analog or pulse signal Default Batch Control Operati...

Page 19: ...age batch control was desired with one relay for control then one would set the assignment of relay 2 assign2 to off rate or grand total The following pages show an illustration of how the above setti...

Page 20: ...PROVU PD6210 Analog Input Batch Controller Instruction Manual 20...

Page 21: ...drive labeled MAINSTAL 3 Double click on the file named MAStart The program will open a few windows and install two programs on your com puter Simply follow the onscreen instructions until you see on...

Page 22: ...rsion factor Enter the conversion factor decimal point and the conversion factor mode Batch mode Automatic or manual batch control T dly Time delay Set time delay for automatic batch restart GT tb Gra...

Page 23: ...tput 1 Program output 1 value e g 4 000 mA Dis 2 Display 2 Program display 2 value Out 2 Output 2 Program output 2 value e g 20 000 mA reset Reset Press Enter to access the Reset menu Rst Hi Reset hig...

Page 24: ...gs are saved to memory only after pressing Enter The display moves to the next menu every time a setting is accepted by pressing Enter Setting Numeric Values The numeric values are set using the Right...

Page 25: ...totalizer and batching features 3 Select the display units tags 4 Decimal point position 5 Program menu for scaling or calibration 6 Display parameter and intensity 7 Relay operation 8 4 20 mA analog...

Page 26: ...features and functions are hidden from the menus Setting the Input Units or Custom Tags units Enter the input unit or custom tag that will be displayed if alternating rate total or grand total and uni...

Page 27: ...librated or scaled to any display value within the range of the controller Additional parameters not needed for most applications are programmed in the Advanced Features menu see ADVANCED OPERATION AN...

Page 28: ...e error message is displayed the controller reverts to input 2 during calibration or scaling and to input 1 during internal calibration allowing the appropriate input signal to be applied or programme...

Page 29: ...lly stopped batch will require a manual start to begin a new batch process Grand Total Reset The grand total can be programmed for manual or automatic reset In the automatic reset mode a programmable...

Page 30: ...min values 5 Display Modbus input 6 Off no display 7 Display rate units 8 Display rate 9 Display total 10 Display grand total 11 Toggle rate and total 12 Toggle rate and grand total 13 Display batch c...

Page 31: ...cess only after alarm condition has cleared e Pump alternation control automatic reset only for level applications unrelated to batch control f Sampling the relay is activated for a user specified tim...

Page 32: ...te for batch control applications but may be used then the controller is in other applications 4 Automatic manual reset when not in a batch process non latching 5 Latching manual reset only when not i...

Page 33: ...for each relay Select on to enable or select off to disable fail safe operation Failsafe mode should not be used on batch relays assigned to total Programming Time Delay The On and Off time delays may...

Page 34: ...ir programmed states when the batch process is started Grand total alarm relays will initialize to their programmed state at controller power up Fail Safe Operation The following table indicates how a...

Page 35: ...as interlock relays To set up the relays for the interlock feature 1 Access the Setup Relay Action menu and set the action to off 2 In the Advanced features User menu program any of the digital input...

Page 36: ...diagram below for more details The utility of this function can of course be expanded beyond sampling and be used whenever a timed relay output closure is required when the batch total interval reach...

Page 37: ...conds Total Relay Sampling Operation When the batch total reaches the set point the relay trips and the sample time starts After the sample time has elapsed the relay resets The cycle repeats every ti...

Page 38: ...same principle applies to the Off time delay Note If Automatic or Manual A mAn reset mode is selected the LED follows the reset point and not the relay state when the relay is acknowledged Scaling th...

Page 39: ...elays and analog output for automatic operation Setting Up the Password pass The Password menu is used for programming three levels of security to prevent unauthorized changes to the programmed parame...

Page 40: ...the menu After exiting the programming mode the controller returns to its password protected condition Disabling Password Protection To disable the password protection access the Password menu and ent...

Page 41: ...tures or make running a batch operation impossible These functions should only be used by users who are sure of their need in their application Advanced Features Menu To simplify the setup process fun...

Page 42: ...onent Set controller for programmable exponent and enter exponent value CutofF Cutoff Set low flow cutoff Count Count Set the totals to count up or down from a set value Tot C Total count Set the batc...

Page 43: ...signal source C CAL Current calibration Calibrating 4 20 mA current input internal reference source used for scaling the input C lo Current low Calibrate low current input e g 4 mA Display Parameter...

Page 44: ...min max display manual alarm reset or total reset will function while a batch is running Advanced Setup and Calibration Multi Point Calibration Scaling The controller is set up at the factory for 2 po...

Page 45: ...nals Rounding is used in addition to the filter function Rounding causes the display to round to the nearest value according to the rounding selected This setting affects the last two digits regardles...

Page 46: ...ransmitter and display flow rate in engineering units Programmable Exponent Linearization Prog E The programmable exponent can be used to linearize the signal from level transmitters in open channel f...

Page 47: ...ned to the rate process variable total grand total maximum or minimum value reached by the rate process one of the set points or the Modbus PV input Analog Output Calibration To perform the analog out...

Page 48: ...Display maximum minimum dis play values on line 2 Ln2 gt Display the grand total on line 2 Ln2 bc Display the batch count on line 2 Contrl Directly access the control menu Display Description disabl...

Page 49: ...w input signal 5 After the display stops flashing a number is displayed with the leftmost digit brighter than the rest The bright digit is the active digit that can be changed by pressing the Up arrow...

Page 50: ...ersion To determine the software version of a controller 1 Go to the Diagnostics menu diAG and press Enter button 2 Press Up arrow button and scroll to Information menu Info 3 Press Enter to access th...

Page 51: ...base GT tb Second Grand total conversion factor T CF 1 000 Grand total reset T rst Manual Relay 1 assignment Asign1 Total Relay 2 assignment Asign2 Total Relay 2 preclose preCls No Relay 3 assignment...

Page 52: ...CutofF 0 000 disabled Maximum output mAx 23 000 mA Minimum output min 1 000 mA Slave ID Address Slave Id 247 Parameter Display Default Setting Baud rate baud 9600 Transmit delay Tr dly 50 ms Parity Pa...

Page 53: ...or calibration Display does not respond to input changes reading a fixed number Check Display assignment it might be displaying max min or set point Display alternates between 1 Hi and a number 2 Lo a...

Page 54: ...PROVU PD6210 Analog Input Batch Controller Instruction Manual 54 This Page Intentionally Left Blank...

Page 55: ...cable technical standards and when applicable or required a European Union notified body certification Standards EN 55022 2003 EN 61000 6 2 2001 EN 61010 1 2001 EN 61326 2006 The standards EN 55022 20...

Page 56: ...0_F SFT039 Ver 4 010 up 03 17 How to Contact Precision Digital For Technical Support please Call 800 610 5239 or 508 655 7300 Fax 508 655 8990 Email support predig com For Sales Support or to place an...

Reviews: