PD6820 Loop-Powered Process Meter Instruction Manual
7
Input
ACCURACY
±0.03% of calibrated span ±1 count,
square root & programmable exponent accuracy range:
10-100% of calibrated span.
ADVANCED
FUNCTION
Linear, square root, or programmable exponent
MULTI-POINT
LINEARIZATION
2 to 32 points
PROGRAMMABLE
EXPONENT
1.0001 to 2.9999
LOW FLOW CUT-
OFF
0-99999 (0 disables cutoff function)
TEMPERATURE
DRIFT
50 PPM/
C from -40 to 75
C ambient
DECIMAL POINT
User selectable decimal point
TOTALIZER
Calculates total based on rate, time base of second, minute,
hour, or day, and field programmable multiplier; stored in non-
volatile memory upon power loss.
TOTALIZER
RESET
User selectable via SafeTouch
®
buttons, time delay, external
contact closure, or protected
CALIBRATION
RANGE
An
Error
message will appear if input 1 and input 2 signals are
too close together.
Input
Range
Minimum Span
Input 1 & Input 2
4-20 mA
0.10 mA
MAXIMUM
VOLTAGE DROP
Without Backlight or with
Externally-Powered (DC
Powered) Backlight
With Loop-Powered Backlight
3.0 VDC @ 20 mA
6.0 VDC @ 20 mA
EQUIVALENT
RESISTANCE
150 Ω @ 20 mA
300 Ω @ 20 mA
EXTERNALLY
POWERED
BACKLIGHT
Voltage Range: Maximum Power
9-36 VDC
9 VDC
12 VDC
24 VDC
36 VDC
0.2 W
0.25 W
0.5 W
0.75 W
INPUT
OVERLOAD
Over current protection to 2 A max.
HART
TRANSPARENCY
Analog input will not interfere with existing HART
communications on the wired 4-20 mA signal