6
Oxygen
Analyzer
Operating Conditions:
Temperature:
50ºF - 113ºF (10ºC - 45ºC)
Altitude:
Sea Level to 8000 feet
Humidity:
0 - 95% non-condensing
Storage Conditions:
Temperature:
5°F - 122°F (-15°C - 50°C)
Humidity:
0 - 95% non-condensing
Mode of Operation:
Continuous
Electrical
Classification:
Internally powered Medical Electrical
equipment
Diverter Fitting:
fits industry standard, 15 mm Tee Adapter
Measurement
Range:
0.0 - 100% Oxygen
Resolution:
0.1 %
Total Accuracy:
± 3.0% Actual Oxygen Level over full
operating temperature range
Drift of
Measurement:
< +/-1% of full scale at constant
temperature, pressure and humidity
Response Time:
90% of final value in less than 12 seconds
at 77°F (25ºC)
Warm-up Time:
Not required
Low Battery
Indication:
Low battery icon displayed
Patient Contact:
Indirect contact via gas passing through
sensor sampling site.
Sensor Specifications
Sensor Type
Galvanic Oxygen Sensor; Precision Medical
PN 504877
Expected Sensor Life
> 1,000,000 O
2
% Hours
Specifications are subject to change without notice.