Precision Medical, Inc. warrants that EASY AIR 15 Compressor (the 

Product) will be free of defects in workmanship and/or material for the 

following period:

One (1) year from shipment

Should any failure to conform to this warranty appear within the 

applicable period, Precision Medical, Inc. shall, upon written 

notification thereof and substantiation that the goods have been 

stored, installed, maintained and operated in accordance with 

Precision Medical, Inc.’s instructions and standard industry practice, 

and that no modifications, substitutions, or alterations have been 

made to the goods, correct such defect by suitable repair or 

replacement at its own expense.

ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.

The representatives of Precision Medical, Inc. or any retailers are not 

authorized to make oral warranties about the merchandise described 

in this contract, and any such statements shall not be relied upon 

and are not part of the contract for sale. Thus, this writing is a final, 

complete and exclusive statement of the terms of that contract.

THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY 

OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR 

OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.

Precision Medical, Inc. shall not under any circumstances be liable for 

special, incidental or consequential damages including but not limited 

to lost profits, lost sales, or injury to person or property. Correction 

of non-conformities as provided above shall constitute fulfillment of 

all liabilities of Precision Medical, Inc. whether based on contract, 

negligence, strict tort or otherwise. Precision Medical, Inc. reserves 

the right to discontinue manufacture of any product or change 

product materials, designs, or specifications without notice. Precision 

Medical, Inc. reserves the right to correct clerical or typographical 

errors without penalty.

Troubleshooting

If the Easy Air Compressor fails to function, consult the Troubleshooting

Guide below. If problem cannot be solved, consult your Provider.

Problem

Probable Cause

Remedy

Compressor  

does not operate

On/Off Switch in “OFF” 

position

Push On/Off Switch to “ON” 

position

Power Cord not connected 

properly

Check Power Cord 

connections

Compressor overheated

Shut Compressor Off, Wait 

15 minutes, Ensure vents are 

clear.

Blown Fuse

Replace Fuse

Power Interruption

Turn off Compressor. When 

Power restored, repeat Start 

Up Procedure.

Trapped back pressure

BEFORE STARTING 

COMPRESSOR. Pull 

“OUT” regulator Knob to 

unlock, then turn Knob fully 

counterclockwise.

Unable to adjust 

pressure/flow

Locked Regulator Knob

Pull Regulator Knob out

Overheating

Poor Ventilation

Allow a minimum of 4 in (10 

cm) clearance on all sides to 

ensure adequate air flow for 

cooling of the Compressor.

Dirty Air Filters

Clean Air Filters

Low pressure/flow

Loose connection at outlet 

fitting

Tighten connection

Improper regulator 

adjustment

Re-adjust regulator

Reviews: