Precision Medical, Inc. warrants that EASY AIR 15 Compressor (the
Product) will be free of defects in workmanship and/or material for the
following period:
One (1) year from shipment
Should any failure to conform to this warranty appear within the
applicable period, Precision Medical, Inc. shall, upon written
notification thereof and substantiation that the goods have been
stored, installed, maintained and operated in accordance with
Precision Medical, Inc.’s instructions and standard industry practice,
and that no modifications, substitutions, or alterations have been
made to the goods, correct such defect by suitable repair or
replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision Medical, Inc. or any retailers are not
authorized to make oral warranties about the merchandise described
in this contract, and any such statements shall not be relied upon
and are not part of the contract for sale. Thus, this writing is a final,
complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY
OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR
OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for
special, incidental or consequential damages including but not limited
to lost profits, lost sales, or injury to person or property. Correction
of non-conformities as provided above shall constitute fulfillment of
all liabilities of Precision Medical, Inc. whether based on contract,
negligence, strict tort or otherwise. Precision Medical, Inc. reserves
the right to discontinue manufacture of any product or change
product materials, designs, or specifications without notice. Precision
Medical, Inc. reserves the right to correct clerical or typographical
errors without penalty.
Troubleshooting
If the Easy Air Compressor fails to function, consult the Troubleshooting
Guide below. If problem cannot be solved, consult your Provider.
Problem
Probable Cause
Remedy
Compressor
does not operate
On/Off Switch in “OFF”
position
Push On/Off Switch to “ON”
position
Power Cord not connected
properly
Check Power Cord
connections
Compressor overheated
Shut Compressor Off, Wait
15 minutes, Ensure vents are
clear.
Blown Fuse
Replace Fuse
Power Interruption
Turn off Compressor. When
Power restored, repeat Start
Up Procedure.
Trapped back pressure
BEFORE STARTING
COMPRESSOR. Pull
“OUT” regulator Knob to
unlock, then turn Knob fully
counterclockwise.
Unable to adjust
pressure/flow
Locked Regulator Knob
Pull Regulator Knob out
Overheating
Poor Ventilation
Allow a minimum of 4 in (10
cm) clearance on all sides to
ensure adequate air flow for
cooling of the Compressor.
Dirty Air Filters
Clean Air Filters
Low pressure/flow
Loose connection at outlet
fitting
Tighten connection
Improper regulator
adjustment
Re-adjust regulator