32
This device has been validated to withstand 5,000 disinfection cycles
using the recommended disinfecting wipe/towelette. The tested number
of cycles is estimated by 1,000 tests per year on each meter and around
3 disinfection cycles per meter per day for 5 years, the expect life of
device.
Stop using the meter if you see any signs of deterioration such as the
LCD display is cracked or cloudy, the buttons no longer function, or the
outer casing is cracked. Please contact the customer service at 1-866-
469-2632 for a replacement meter if any of the signs of deterioration are
noticed.
The meter should be replaced after the validated number of disinfection
cycles or the warranty period, which ever comes
first.
NOTE:
• Do
NOT
clean or disinfect the meter while performing a test.
• If the meter is being operated by a second person, the meter and lancing
device should be decontaminated prior to handling by an assistant.
• Do
NOT
allow cleaning or disinfecting solution to get in the test slot, battery
compartment, or strip-ejection button.
• If you do get moisture in the test strip slot, wipe it away with the corner of a
tissue.
• Always dry the meter thoroughly before using it.
• Do not spray the meter directly with cleaning solutions especially those
containing water (i.e. soapy water), as this could cause the solution to enter
the case and damage the electronic components or circuitry.