XTR-PX2-C0006-IN004 Ceph option
2-12
2.15 Adjusting the Attaching Angle of CEPHALO ASSY (1)
(1) Set up a level on the CEPHALO ARM (shown in the picture below).
(2) Adjust the CEPHALO ARM while turning the adjustment screw (CP8X30-CM1) of CEPHALO ARM BASE
so that the level display turns to
“0.0±0.1”.
(3) Tighten the fixing screws of CEPHALO ARM BASE (SCBST8-22) (x3).
(4) Check that the level display has become
“0±0.1°” after fixing it.
2.16 Adjusting the Attaching Angle of CEPHALO ASSY (2)
(1) Loosen the fixing screws. (
mark)
(2) Set up a level at Level setup place [1].
(3) Adjust the CEPHALO ARM while turning the adjustment screw [1] (ULTRA-LOW HEAD CP5X10-BK:
Screw bag G) so that the level display turns to
“0.0±0.1°”.
(4) The same as step (2), set up another level at Level setup place [2].
(5) Adjust the CEPHALO ARM while turning the adjustment screw [2] (ULTRA-LOW HEAD CP5X10-BK:
Screw bag G) so that the level display turns to
“0.0±0.1°”.
(6) Tighten the fixing screws.
(7) Check that the level displays at both setup place of [1] and [2] are
“0.0±0.1°”.
(This is not a problem as the adjustment screw is mounted as is)
Set up a level on the CEPHALO ARM
Adjustment screw
(CP8X30-CM1)
Fixing screws
(SCBST8-22) (x3)
(2) Level setup place [1]
(4) Level setup place [2]
(1) Fixing screw
(3) Adjustment screw [1]
(5) Adjustment screw [2]
Setup example of the level
Adjustment screw [1]