V I I . C A R E A N D M A I N T E N A N C E
10. Place the batteries back onto the power base.
11. Plug in the quick-disconnects.
12. Reinstall the shroud.
13. Reinstall the seat.
14. Reinstall the controller communications cable.
15. Charge the batteries.
When to See Your Authorized Provider/Specialist for Service
The following symptoms could indicate a serious problem with your power chair. If necessary, contact your autho-
rized provider/specialist. When calling, have the model number, serial number, nature of the problem, and the trouble
code if available.
!
Motor noise
!
Frayed harnesses
!
Cracked or broken connectors
!
Uneven wear on any of the tires
!
Jerky motion
!
Pulling to one side
!
Bent or broken wheel assemblies
!
Does not power up
!
Powers up, but does not move
Corrective Maintenance
If the battery condition meter does not light up when you turn on the power:
!
Check the harness connections. Make sure they are tight.
!
Check the circuit breaker. Reset it if necessary.
!
Check the battery connections.
If the above conditions prove normal, you can load test the batteries with a battery load tester. These testers are available
at automotive parts stores. Disconnect both batteries before load testing and follow the directions that come with the load
tester. If either one of the batteries fails the load test, replace both of them. If your power chair still does not power up,
contact your authorized provider/specialist.