Menu: SET (Setup)
SET 6
SET 6
If you select „PTSA-/ Fluorescein-adapter“:
Select a 24mm glass vial for future PTSA and Fluorescein tests.
Fill vial with 10ml of distilled water.
USE PIPETTE TO ENSURE 10ml!
Place vial into the PrimeLab. Do not close it!
Place PTSA-/Fluorescein-adapter on top of the open cuvette.
Press TEST.
Remove vial from PrimeLab.
Empty vial.
Fill vial with 10ml of reference-solution „500 ppb PTSA“ / „100 ppb
Fluorescein“.
USE PIPETTE!
Place vial into the PrimeLab. Do not close it!
Place PTSA-/Turbidity-adapter on top of the open cuvette.
Press TEST. Calibation was sucessful!
Please use ONLY THIS VIAL for future PTSA/Flurescein tests!
Error Messages
During calibration and / or during the measurement, the following error
messages might be displayed on the PrimeLab:
Adapter not recognized - Possible causes:
Ÿ
Battery depleted
Ÿ
Adapter incorrectly (eg diagonally) placed
Ÿ
Wrong adapter used (eg PTSA instead Turbidity)
Ÿ
Adapter-lense (bottom) or optical path in the sample chamber
Ÿ
(PrimeLab) dirty or wet
Measurement failed - Possible causes:
Ÿ
Battery depleted
Ÿ
Adapter incorrectly (eg diagonally) placed
Ÿ
The identified water sample does not match the turbidity curve
Ÿ
Wrong adapter used (eg PTSA instead of Turbidity)
Ÿ
Adapter-lense (bottom) or optical path in the sample chamber
Ÿ
(PrimeLab) dirty or wet
If the message reads “Calibration: ERROR!” or
“LED problem. Please check!” please abort the procedure and check that
there is no cuvette in the sampling chamber and that the sampling
chamber and especially the two “windows” on the sensor / LED level are
clean (see also “Cleaning the device” on page A 3).
Restart the calibration procedure.
If the device fails to calibrate after a second attempt it must be returned to
the manufacturer for inspection.
Continued...
Calibration
The following refers to calibration of ID 111/113 (PTSA/Fluorescein):
For ID 111 and ID 113 it is essential to always use the same vial for calibration as well as
for testing as adapter shines from above and each vial has different bottom
Due to the above it is essential to always use the same vial for testing which has
been used for calibration!
4
4
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...