122
122
0 - 10 mg/l
2+/3+
(Fe
)
Name on device: 29-Iron-MR-pow
(29)
Iron (MR)
Continued...
PL Iron MR 1
(PLpow20IronMR1)
Powder
10 ml
24 mm
Filter
0.45µm
(*)
(*)
(*)
Syringe
10
9
8
7
6
5
4
3
2
1
ml
Filter paper Filter adapter
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
Select either
A) "dissolved iron" (*) or
B) "total iron"
A) Measurement procedure for “dissolved iron”
1) Unscrew the two halves of a clean and free of residues filter adapter*.
Insert 25mm filter paper* (0.45 micron).
2) Close the two halves of the filter adapter with the filter paper inside.
Make sure the seal ring is seated correctly.
3) Fill the 20ml filter syringe* (clean and free of residues) with 14ml of
sample water.
4) Screw the filter adapter prepared in steps (1) and (2) onto the syringe
point and empty it to the 10ml level.
5) Push the remaining 10ml in the syringe through the filter adapter into a
clean 24mm cuvette.
6) Replace the screw lid on the cuvette and insert it into the PrimeLab
sampling chamber. Align the arrow on the front of the cuvette. Press
ZERO.
7) Remove the cuvette from the chamber and unscrew the lid.
4
2
3
4
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 6)
-> Use last ZERO
= continue with step 8)
1
2
3
4
Powder
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...