141
141
Note(s)
->
Result is displayed as “cfu/l” which is related to filtration of 1 liter of your sample.
->
If less than 1 liter is filtered, e.g. only 500ml, results needs to be read as “cfu/..” (e.g. cfu/500ml).
->
Once reagents are received, kit MUST be stored b2°C and +8°C, preferrably at +4°C.
->
Expiry date of the reagents is 3 months from production date on.
->
Avoid contact with eyes. Wear protective gloves.
->
Certain isolates cannot be detected below 106 cfu.
->
Disposal of product according to local regulations. Products are stable and unlikely to react
in a hazardous manner under normal conditions of use.
->
Do NOT re-use small 1ml vials (LG-CB).
->
Leave at least 12 cm space between multiple LG-MH (magnetic holders).
->
Reagents are supplied in excess. Do NOT re-use any leftover amounts of reagents.
->
When emptying cuvettes LG-MHCB, always do so to the BACK and never in front (magnet)!
->
You need to properly follow test procedure to avoid mis-readings.
->
Once lids of LG-MHCB are removed and to discarded, do NOT use them for any of the
following test steps.
->
If you do more than 1 test at the same time, only one blanc/ZERO vial is needed.
->
We propose to use LG-MP4 automatic agitator plate to place up to 20 LG-MHCB cuvettes
in case you do multiple tests at one time.
->
Measurement has to be performed immediately after the last step (countdown), as the color
reaction might continue.
->
Leaving reagents at room temperature for 30 minutes before starting the test is essential.
->
When using larger units of reagents, immediately restore in fridge after use.
->
Depending on the water quality of the test water, the pre-filter have to be changed during
the filter process, if it is too dirty.
Continued...
Name on device: 147-Legionella-liq
(147)
Legionella
Liquid
60 - 1000000 cfu/test
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...